Johns Hopkins University School of Medicine, Baltimore, MD, USA.
Eye (Lond). 2012 Mar;26(3):454-62. doi: 10.1038/eye.2011.293. Epub 2011 Dec 2.
To report macular thickness values in normal eyes and eyes with diabetic macular edema (DME) using time-domain (TD) and spectral-domain (SD) optical coherence tomography (OCT), and to derive a conversion equation.
The index study was a prospective investigation conducted on 80 eyes from 40 normal subjects and 130 eyes from 118 patients with DME seen in our clinic. Retinal thickness values from the central 1 mm of the macula and surrounding four ETDRS subfields were acquired using TD-OCT (Stratus OCT) and SD-OCT (SPECTRALIS HRA+OCT). Measurements of the central (C) subfield from both devices were used to derive a conversion equation. The equation was used to predict SD-OCT values using measurements from TD-OCT. Agreement between predicted and actual SD-OCT measurements was assessed.
In normal eyes, the mean difference between TD-OCT and SD-OCT measurements of the C subfield was 76 μm (CI(95)=74 and 77, respectively). The conversion equation, y=1.029x+72.49, was derived. In eyes with DME, using the equation, SPECTRALIS-predicted values were 5% higher than actual measurements, with 95% of predicted values falling within 9% of the actual measurements. Relocating SD-OCT grids to match the location on TD-OCT resulted in predicted values falling within 7% of actual measurements.
The percent difference between actual thickness measurements from SPECTRALIS and predicted thickness measurements, using the conversion equation, was within reported limits of repeatability of Stratus in eyes with DME. Our equation may help correlate OCT values from both devices in standard care and clinical trials for DME.
利用时域(TD)和频域(SD)光学相干断层扫描(OCT)报告正常眼和糖尿病性黄斑水肿(DME)眼中的黄斑厚度值,并推导出转换方程。
本研究为前瞻性研究,纳入 40 名正常受试者的 80 只眼和我们诊所就诊的 118 例 DME 患者的 130 只眼。使用 TD-OCT(Stratus OCT)和 SD-OCT(SPECTRALIS HRA+OCT)获取黄斑中央 1mm 及周围四个 ETDRS 亚区的视网膜厚度值。使用两种设备的中央(C)亚区的测量值推导出转换方程。该方程用于根据 TD-OCT 的测量值预测 SD-OCT 值。评估预测和实际 SD-OCT 测量值之间的一致性。
在正常眼中,C 亚区的 TD-OCT 和 SD-OCT 测量值之间的平均差异为 76μm(CI(95)分别为 74 和 77μm)。推导出转换方程 y=1.029x+72.49。在 DME 眼中,使用该方程,SPECTRALIS 预测值比实际测量值高 5%,95%的预测值落在实际测量值的 9%以内。将 SD-OCT 网格重新定位以匹配 TD-OCT 的位置,使预测值落在实际测量值的 7%以内。
使用转换方程,SPECTRALIS 的实际厚度测量值与预测厚度测量值之间的差异百分比在 DME 眼中 Stratus 报告的可重复性范围内。我们的方程可帮助在标准护理和 DME 的临床试验中关联两种设备的 OCT 值。