Sayanagi Kaori, Sharma Sumit, Kaiser Peter K
Cole Eye Institute, Cleveland Clinic Foundation, Cleveland, Ohio 44195, USA.
Am J Ophthalmol. 2009 Sep;148(3):431-8. doi: 10.1016/j.ajo.2009.04.008. Epub 2009 Jun 3.
To compare retinal thickness (RT) measurements between traditional 6 radial line scans and three-dimensional (3D) scans using spectral-domain optical coherence tomography (SD-OCT).
Prospective, consecutive case series.
Twenty eyes of 17 patients with macular diseases and 8 healthy subjects were scanned using 2 different methods (3D scan and 6 radial line scan) using 2 different SD-OCT devices (Topcon OCT-1000 [Topcon Inc, Paramus, New Jersey, USA] and Canon SPOCT [(Canon/Optopol Inc, Depew, New York, USA]) by an experienced OCT operator. In 16 eyes, the data of TD-OCT were also obtained for comparison. The RT in 9 regions based on Early Treatment Diabetic Retinopathy Study areas were analyzed and compared. Bland-Altman plots were used to evaluate agreement.
In the eyes with macular diseases, there was no significant difference in the RT between 3D and 6 radial line scans except the center subfield (P = .011) on the Topcon OCT-1000, and inner inferior (P = .031) and outer nasal (P = .014) areas on the Canon SPOCT. In the healthy subjects, there was no significant difference between the different scans on both SD-OCT devices. The 95% limit of subjects was 3.2 to 30.6 microm and 16.7 to 28.3 microm in the eye with macular diseases on OCT-1000 and SPOCT, respectively, and 6.7 to 13.6 microm and 10.8 to 32.3 microm in healthy subjects on OCT-1000 and SPOCT, respectively.
Preliminary results suggest the RT obtained by the 2 different scan patterns, 3D scan and 6 radial line scan, did not differ significantly on SD-OCT.
使用光谱域光学相干断层扫描(SD-OCT)比较传统的6条径向线扫描与三维(3D)扫描之间的视网膜厚度(RT)测量值。
前瞻性、连续病例系列。
由一名经验丰富的OCT操作人员使用2种不同的SD-OCT设备(拓普康OCT-1000 [美国新泽西州帕拉默斯市拓普康公司] 和佳能SPOCT [美国纽约州迪皮尤市佳能/奥普托波尔公司]),采用2种不同方法(3D扫描和6条径向线扫描)对17例黄斑疾病患者的20只眼和8名健康受试者进行扫描。在16只眼中,还获取了时域光学相干断层扫描(TD-OCT)数据用于比较。分析并比较基于糖尿病视网膜病变早期治疗研究区域的9个区域的视网膜厚度。采用Bland-Altman图评估一致性。
在黄斑疾病患者的眼中,除了拓普康OCT-1000上的中心子区域(P = 0.011),以及佳能SPOCT上的内下区域(P = 0.031)和外鼻侧区域(P = 0.014)外,3D扫描和6条径向线扫描之间的视网膜厚度无显著差异。在健康受试者中,两种SD-OCT设备上不同扫描之间均无显著差异。在黄斑疾病患者眼中,OCT-1000和SPOCT上受试者的95%界限分别为3.2至30.6微米和16.7至28.3微米;在健康受试者中,OCT-1000和SPOCT上受试者的95%界限分别为6.7至13.6微米和10.8至32.3微米。
初步结果表明,在SD-OCT上,通过3D扫描和6条径向线扫描这两种不同扫描模式获得的视网膜厚度无显著差异。