Measurement of hepatic lipid: high-speed T2-corrected multiecho acquisition at 1H MR spectroscopy--a rapid and accurate technique.

作者信息

Pineda Nashiely, Sharma Puneet, Xu Qin, Hu Xiaoping, Vos Miriam, Martin Diego R

机构信息

Department of Biomedical Engineering, Georgia Institute of Technology, Atlanta, GA, USA.

出版信息

Radiology. 2009 Aug;252(2):568-76. doi: 10.1148/radiol.2523082084. Epub 2009 Jun 22.

Abstract

PURPOSE

To evaluate the feasibility, accuracy, and reproducibility of a fast breath-hold magnetic resonance (MR) spectroscopic method for T2-corrected hepatic lipid measurement in phantoms and in humans.

MATERIALS AND METHODS

All experiments were institutional review board approved and HIPAA compliant; informed consent was obtained from all subjects. The 15-second breath-hold high-speed T2-corrected multiecho (HISTO) MR spectroscopic technique was developed to acquire multiple echoes in a single acquisition, which enables the quantification of water and lipid T2, and subsequently to provide a corrected measure of hepatic lipid fraction. The accuracy of T2-corrected MR spectroscopy was evaluated in eight lipid phantoms doped with iron to simulate variable T2 effects. The mean absolute error of the HISTO technique with the known lipid amounts, as well as with uncorrected MR spectroscopic measures, was evaluated. The HISTO sequence was performed in 25 male subjects (mean age, 23.0 years +/- 19.2 [standard deviation]) to evaluate measurement bias with conventional, uncorrected MR spectroscopy. Three additional male subjects (mean age, 30.0 years +/- 1.0) were examined to assess reproducibility by using analysis of variance testing within subject and between separate imaging sessions.

RESULTS

The absolute error in quantifying lipid fraction by using iron-doped lipid phantoms was less than 11% for the HISTO technique, compared with more than 50% for uncorrected MR spectroscopy. In the 25 human subjects, hepatic lipid measured by using HISTO differed significantly from that by using uncorrected MR spectroscopic methods by 5.1% +/- 2.6. Analysis of variance of three separate imaging sessions with the HISTO technique indicated no significant variance (P = .13) in three subjects.

CONCLUSION

HISTO is an accurate, reproducible MR spectroscopic sequence for quantifying hepatic lipid noninvasively. Evidence has shown this method to be feasible in vivo for clinical use.

摘要

目的

评估一种快速屏气磁共振(MR)波谱法在体模和人体中进行T2校正肝脂质测量的可行性、准确性和可重复性。

材料与方法

所有实验均经机构审查委员会批准并符合健康保险流通与责任法案(HIPAA)规定;所有受试者均获得知情同意。开发了15秒屏气高速T2校正多回波(HISTO)MR波谱技术,以在一次采集过程中获取多个回波,从而能够对水和脂质的T2进行定量,并随后提供肝脂质分数的校正测量值。在八个掺杂铁的脂质体模中评估T2校正MR波谱的准确性,以模拟不同的T2效应。评估了HISTO技术在已知脂质含量情况下以及未校正MR波谱测量情况下的平均绝对误差。对25名男性受试者(平均年龄23.0岁±19.2[标准差])进行HISTO序列检查,以评估与传统的、未校正的MR波谱相比的测量偏差。另外三名男性受试者(平均年龄30.0岁±1.0)接受检查,通过受试者内和不同成像 session 之间的方差分析来评估可重复性。

结果

对于HISTO技术,使用掺杂铁的脂质体模定量脂质分数时的绝对误差小于11%,而未校正的MR波谱法超过50%。在25名人类受试者中,使用HISTO测量的肝脂质与使用未校正MR波谱法测量的结果相比,差异显著,为5.1%±2.6。对三名受试者进行的三次不同成像 session 的HISTO技术方差分析表明无显著差异(P = 0.13)。

结论

HISTO是一种准确、可重复的MR波谱序列,用于无创定量肝脂质。有证据表明该方法在体内临床应用中是可行的。