Suppr超能文献

两种基于频率的电子根尖定位仪工作长度测定的体内比较

An in vivo comparison of working length determination of two frequency-based electronic apex locators.

作者信息

Pascon E A, Marrelli M, Congi O, Ciancio R, Miceli F, Versiani M A

机构信息

Calabrodental Centro Odontoiatrico, Crotone, Italy.

出版信息

Int Endod J. 2009 Nov;42(11):1026-31. doi: 10.1111/j.1365-2591.2009.01615.x.

Abstract

AIM

To compare in vivo the accuracy of two electronic apex locators (EALs) by means of digital radiographic imaging system.

METHODOLOGY

Electronic working lengths of 831 canals were determined with the DentaPort ZX and Raypex 5 apex locators and confirmed radiographically. The radiographic images acquired with the aid of a digital radiographic imaging system (VisualiX eHD; Gendex Dental Systems, Des Plaines, IL, USA) were blindly analysed by two independent evaluators. The distance between the file tip and the radiographic apex was measured using dedicated software (VixWin Pro, Gendex Dental Systems, Des Plaines, IL, USA) and the mean distance achieved between different tooth type and EALs were compared statistically. Statistical analyses were performed using the t-test for independent samples and one-way anova with the null hypothesis set as 5%. Positive or negative values were recorded when the file tip was detected beyond or short of the radiographic apex, respectively.

RESULTS

The mean distance between file tip and radiographic apex were -1.08 +/- 0.73 and -1.0 +/- 0.67 mm considering DentaPort ZX and Raypex 5 groups, respectively, with no significant differences (P > 0.05). No statistically significant differences were found amongst the same tooth type when comparing both groups (P > 0.05) or amongst different teeth type in the same group (P > 0.05).

CONCLUSIONS

Within the limitations of this in vivo study, the DentaPort ZX and Raypex 5 were similar in terms of accuracy.

摘要

目的

通过数字放射成像系统在体内比较两种电子根尖定位仪(EALs)的准确性。

方法

使用DentaPort ZX和Raypex 5根尖定位仪确定831个根管的电子工作长度,并通过放射成像进行确认。由两名独立评估人员对借助数字放射成像系统(VisualiX eHD;美国伊利诺伊州德斯普兰斯的Gendex Dental Systems公司)获取的放射图像进行盲法分析。使用专用软件(VixWin Pro,美国伊利诺伊州德斯普兰斯的Gendex Dental Systems公司)测量锉尖与放射学根尖之间的距离,并对不同牙型和EALs之间的平均距离进行统计学比较。采用独立样本t检验和单因素方差分析进行统计分析,原假设设定为5%。当锉尖位于放射学根尖之外或不足时,分别记录正值或负值。

结果

考虑DentaPort ZX组和Raypex 5组,锉尖与放射学根尖之间的平均距离分别为-1.08±0.73和-1.0±0.67mm,无显著差异(P>0.05)。比较两组时,同一牙型之间(P>0.05)或同一组不同牙型之间(P>0.05)均未发现统计学显著差异。

结论

在本体内研究的局限性内,DentaPort ZX和Raypex 5在准确性方面相似。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验