National Reference Center for Keratoconus, Department of Ophthalmology, University Hospital Pellegrin, Bordeaux, France.
Invest Ophthalmol Vis Sci. 2010 May;51(5):2403-10. doi: 10.1167/iovs.09-3689. Epub 2009 Nov 11.
To evaluate the performance of the Ocular Response Analyzer (ORA) in the screening of forme fruste keratoconus (FFKc).
A retrospective comparative study was conducted involving 180 eyes. ORA preoperative data were analyzed for 125 normal control eyes (64 patients) undergoing laser in situ keratomileusis (LASIK) without corneal ectasia after 24 months of follow-up and 55 case eyes with unilateral keratoconus from a database (BCVA of 1.0, KISA index <60%). All eyes were matched in four groups of central corneal thickness (CCT): group 1, <500 microm; group 2, 500 to 539 microm; group 3, 540 to 579 microm; and group 4 >580 microm. Corneal hysteresis (CH), the corneal resistance factor (CRF), the air pressure curve, and the infrared signal were compared between FFKc and normal eyes in each group.
The mean CH was 9.1 +/- 1.8 mm Hg for FFKc and 10.3 +/- 1.9 mm Hg for control eyes (P < 0.001), and the mean CRF was 9.2 +/- 1.8 and 11.1 +/- 2 mm Hg (P < 0.001), respectively. Sensitivity in each group was as follows: group 1, CH < 9.5 mm Hg (91%) and CRF < 9.5 mm Hg (81%); group 2, CH < 10.5 mm Hg (91%) and CRF < 10 mm Hg (87%); group 3, CH < 11.5 mm Hg (79%) and CRF < 11 mm Hg (74%); group 4 had two cases of FFKc, and the difference was not significant. Air pressure levels at inward and outward applanation and the maximum air pressure level were significantly lower and shorter in time in FFKc (P < 0.001), whereas the shape of the infrared signal was more variable.
The ORA provides additional information in the screening of FFKc, with an accurate analysis of the corneal biomechanical properties according to CCT, air pressure, and infrared curves.
评估眼反应分析仪(ORA)在筛查非典型性圆锥角膜(FFKc)中的性能。
这是一项回顾性对比研究,共纳入 180 只眼。对数据库中 55 只单侧圆锥角膜患眼(BCVA 为 1.0,KISA 指数<60%)和 125 只行 LASIK 术后 24 个月无角膜膨隆的正常对照眼(64 例)的术前 ORA 数据进行分析。所有眼均按中央角膜厚度(CCT)分为 4 组:组 1,<500μm;组 2,500-539μm;组 3,540-579μm;组 4,>580μm。比较每组中 FFKc 眼与正常眼的角膜滞后量(CH)、角膜阻力因子(CRF)、眼压曲线和红外信号。
FFKc 眼的平均 CH 为 9.1±1.8mmHg,正常眼为 10.3±1.9mmHg(P<0.001),平均 CRF 分别为 9.2±1.8mmHg 和 11.1±2mmHg(P<0.001)。每组的敏感性如下:组 1,CH<9.5mmHg(91%)和 CRF<9.5mmHg(81%);组 2,CH<10.5mmHg(91%)和 CRF<10mmHg(87%);组 3,CH<11.5mmHg(79%)和 CRF<11mmHg(74%);组 4 有 2 例 FFKc,差异无统计学意义。FFKc 眼的内压和外压水平以及最大眼压水平明显较低,用时较短(P<0.001),而红外信号的形状则更为多变。
ORA 根据 CCT、眼压和红外曲线对角膜生物力学特性进行准确分析,为 FFKc 的筛查提供了额外信息。