超声角膜测厚术与CASIA2(眼前节光学相干断层扫描)在测量中央角膜厚度方面的比较。

A Comparison Between Ultrasound Pachymetry and CASIA2 (Anterior-Segment Optical Coherence Tomography) in the Measurement of Central Corneal Thickness.

作者信息

Baxter Joe, Atwan Nadeem

机构信息

Ophthalmology, Sheffield Teaching Hospitals NHS Foundation Trust, Sheffield, GBR.

出版信息

Cureus. 2023 Jun 3;15(6):e39921. doi: 10.7759/cureus.39921. eCollection 2023 Jun.

Abstract

Background and objective Due in part to its effect on intraocular pressure (IOP) measurements, the assessment of central corneal thickness (CCT) is recognized as an essential part of the initial glaucoma assessment. The most widely utilized clinical technique to measure CCT is ultrasound pachymetry (USP). In recent years, many dedicated anterior-segment optical coherence tomography scanners (AS-OCTs) have been developed. Previous studies have compared CCT measurements between USP and various AS-OCTs. This study aimed to assess the degree of agreement between USP and CASIA2 (Tomey Corporation, Nagoya, Japan), a second-generation swept-source AS-OCT developed in Japan. Methodology The data on CCT screening measurements of 156 eyes (88 patients) performed over a period of three months, from January to March 2020, on glaucoma patients attending the Royal Hallamshire Hospital (RHH) in Sheffield, UK were collected retrospectively and statistically analyzed. Results The average age of the 88 patients included in the study was 66 years (range: 20-86 years). Our findings show that when compared to CASIA2 measurements, USP measurement of the CCT resulted in significantly thicker values (paired t-test: t=23.15,p<2.2 x 10). The average difference between the two methods was 19.98 ± 10.78 μm. It is hypothesized that this difference may be due in part to inaccurate probe placement during ultrasound probe measurement, resulting in thicker CCT values. Conclusion The observed difference may be clinically significant as it could induce clinical discrepancy in terms of perceived glaucoma risk in patients. Therefore, USP and CASIA2 should not be used interchangeably, and clinicians should take into account the significant difference between these methods.

摘要

背景与目的 部分由于其对眼压(IOP)测量的影响,中央角膜厚度(CCT)的评估被认为是青光眼初始评估的重要组成部分。测量CCT最广泛使用的临床技术是超声角膜测厚法(USP)。近年来,已开发出许多专用的眼前节光学相干断层扫描设备(AS - OCT)。先前的研究比较了USP与各种AS - OCT之间的CCT测量结果。本研究旨在评估USP与日本开发的第二代扫频源AS - OCT CASIA2(日本名古屋Tomey公司)之间的一致性程度。

方法 回顾性收集并统计分析了2020年1月至3月期间在英国谢菲尔德皇家哈勒姆郡医院(RHH)对青光眼患者进行的为期三个月的156只眼(88例患者)的CCT筛查测量数据。

结果 纳入研究的88例患者的平均年龄为66岁(范围:20 - 86岁)。我们的研究结果表明,与CASIA2测量结果相比,USP测量的CCT值明显更厚(配对t检验:t = 23.15,p < 2.2×10)。两种方法的平均差异为19.98±10.78μm。据推测,这种差异可能部分是由于超声探头测量期间探头放置不准确,导致CCT值更厚。

结论 观察到的差异可能具有临床意义,因为它可能在患者青光眼风险感知方面引起临床差异。因此,USP和CASIA2不应互换使用,临床医生应考虑这些方法之间的显著差异。