Department of Nuclear and Radiological Engineering, University of Florida, 202 Nuclear Science Building, Gainesville, Florida 32611-8300, USA.
Med Phys. 2009 Oct;36(10):4495-503. doi: 10.1118/1.3218767.
The authors have developed a quantitative calibration method for a multileaf collimator (MLC) which measures individual leaf positions relative to the MLC backup jaw on an Elekta Synergy linear accelerator.
The method utilizes a commercially available two-axis detector array (Profiler 2; Sun Nuclear Corporation, Melbourne, FL). To calibrate the MLC bank, its backup jaw is positioned at the central axis and the opposing jaw is retracted to create a half-beam configuration. The position of the backup jaws field edge is then measured with the array to obtain what is termed the radiation defined reference line. The positions of the individual leaf ends relative to this reference line are then inferred by the detector response in the leaf end penumbra. Iteratively adjusting and remeasuring the leaf end positions to within specifications completes the calibration. Using the backup jaw as a reference for the leaf end positions is based on three assumptions: (1) The leading edge of an MLC leaf bank is parallel to its backup jaw's leading edge, (2) the backup jaw position is reproducible, and (3) the measured radiation field edge created by each leaf end is representative of that leaf's position. Data from an electronic portal imaging device (EPID) were used in a similar analysis to check the results obtained with the array.
The relative leaf end positions measured with the array differed from those measured with the EPID by an average of 0.11+/-0.09 mm per leaf. The maximum leaf positional change measured with the Profiler 2 over a 3 month period was 0.51 mm. A leaf positional accuracy of +/-0.4 mm is easily attainable through the iterative calibration process. The method requires an average of 40 min to measure both leaf banks.
This work demonstrates that the Profiler 2 is an effective tool for efficient and quantitative MLC quality assurance and calibration.
作者开发了一种用于 Elekta Synergy 线性加速器的多叶准直器(MLC)的定量校准方法,该方法测量叶片相对于 MLC 挡块的单个叶片位置。
该方法使用商业上可获得的二维探测器阵列(Profiler 2;Sun Nuclear Corporation,墨尔本,FL)。为了校准 MLC 库,将其挡块定位在中心轴上,并缩回相对的挡块以形成半束配置。然后使用阵列测量挡块场边缘的位置,以获得所谓的辐射定义参考线。然后通过探测器在叶片末端半影中的响应推断单个叶片末端相对于该参考线的位置。通过在规格内反复调整和重新测量叶片末端位置来完成校准。基于三个假设,使用挡块作为叶片末端位置的参考:(1)MLC 叶片库的前缘与其挡块的前缘平行,(2)挡块位置可重复,(3)每个叶片末端创建的测量辐射场边缘代表该叶片的位置。使用电子门户成像设备(EPID)中的数据进行类似的分析,以检查与阵列获得的结果。
与 EPID 相比,阵列测量的叶片末端位置平均相差 0.11±0.09 毫米/叶。Profiler 2 在 3 个月期间测量的最大叶片位置变化为 0.51 毫米。通过迭代校准过程,可以轻松实现叶片位置精度为 +/-0.4 毫米。该方法平均需要 40 分钟来测量两个叶片库。
这项工作表明,Profiler 2 是一种用于高效、定量 MLC 质量保证和校准的有效工具。