Cardiac magnetic resonance imaging using radial k-space sampling and self-calibrated partial parallel reconstruction.

机构信息

Department of Radiology, Northwestern University, Chicago, IL, USA.

出版信息

Magn Reson Imaging. 2010 May;28(4):495-506. doi: 10.1016/j.mri.2009.12.018. Epub 2010 Jan 12.

Abstract

Radial sampling has been demonstrated to be potentially useful in cardiac magnetic resonance imaging because it is less susceptible to motion than Cartesian sampling. Nevertheless, its capability of imaging acceleration remains limited by undersampling-induced streaking artifacts. In this study, a self-calibrated reconstruction method was developed to suppress streaking artifacts for highly accelerated parallel radial acquisitions in cardiac magnetic resonance imaging. Two- (2D) and three-dimensional (3D) radial k-space data were collected from a phantom and healthy volunteers. Images reconstructed using the proposed method and the conventional regridding method were compared based on statistical analysis on a four-point scale imaging scoring. It was demonstrated that the proposed method can effectively remove undersampling streaking artifacts and significantly improve image quality (P<.05). With the use of the proposed method, image score (1-4, 1=poor, 2=good, 3=very good, 4=excellent) was improved from 2.14 to 3.34 with the use of an undersampling factor of 4 and from 1.09 to 2.5 with the use of an undersampling factor of 8. Our study demonstrates that the proposed reconstruction method is effective for highly accelerated cardiac imaging applications using parallel radial acquisitions without calibration data.

摘要

径向采样在心脏磁共振成像中被证明是有用的,因为它比笛卡尔采样更不容易受到运动的影响。然而,其成像加速的能力仍然受到欠采样引起的条纹伪影的限制。在这项研究中,开发了一种自校准重建方法,以抑制心脏磁共振成像中高度加速的并行径向采集的条纹伪影。在体模和健康志愿者中采集了二维 (2D) 和三维 (3D) 径向 k 空间数据。基于四点成像评分的统计分析,比较了使用所提出的方法和传统重排方法重建的图像。结果表明,该方法可以有效地去除欠采样条纹伪影,显著提高图像质量 (P<.05)。使用所提出的方法,在使用 4 倍欠采样因子和 8 倍欠采样因子的情况下,图像评分(1-4,1=差,2=好,3=很好,4=优秀)从 2.14 提高到 3.34。我们的研究表明,该重建方法在没有校准数据的情况下,对于使用并行径向采集的高度加速心脏成像应用是有效的。

相似文献

7
Self-calibration method for radial GRAPPA/k-t GRAPPA.径向GRAPPA/k-t GRAPPA的自校准方法
Magn Reson Med. 2007 Jun;57(6):1075-85. doi: 10.1002/mrm.21233.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验