Department of Otolaryngology, Kaohsiung Municipal Hsaio-Kang Hospital, Kaohsiung Medical University, Kaohsiung City, Taiwan.
Head Neck. 2010 Oct;32(10):1295-301. doi: 10.1002/hed.21324.
Intraoperative neuromonitoring (IONM) of the recurrent laryngeal nerve (RLN) has recently been more frequently applied in thyroid surgery. However, concerns have been raised regarding the safety and optimal intensity of electrical nerve stimulation.
Eight piglets were enrolled, and electrically evoked electromyography (EMG) was recorded from the vocalis muscles via endotracheal surface electrodes. The baseline EMG was measured and continuous pulsatile stimulations were performed on the vagus nerve and RLN for 10 minutes. Changes of EMG waveform and cardiopulmonary status were analyzed.
A dose-response curve existed with increasing EMG amplitude as stimulating current was increased, with maximum amplitude elicited on vagal and RLN stimulation at <1 mA. No obvious EMG changes and untoward cardiopulmonary effects were observed after the stimulation.
Electrical stimulation is safe during IONM in this porcine model. Minimal current that required generating the maximal evoked EMG, approximately 1 mA in this study, can be selected to minimize the risk of nerve damage and cardiopulmonary effects.
术中喉返神经(RLN)监测(IONM)最近在甲状腺手术中被更频繁地应用。然而,人们对电神经刺激的安全性和最佳强度提出了担忧。
纳入 8 头小猪,通过气管内表面电极从杓状肌记录电诱发肌电图(EMG)。测量基础 EMG,并对迷走神经和 RLN 进行 10 分钟的连续脉冲刺激。分析 EMG 波形和心肺状态的变化。
存在剂量-反应曲线,随着刺激电流的增加,EMG 幅度增加,迷走神经和 RLN 刺激的最大幅度在<1 mA。刺激后无明显的 EMG 变化和不良心肺作用。
在本猪模型中,IONM 期间电刺激是安全的。在这项研究中,产生最大诱发 EMG 所需的最小电流约为 1 mA,可以选择最小化神经损伤和心肺作用的风险。