一种新型双极射频消融装置在心脏跳动时进行房颤消融的疗效:一项长期猪实验研究

Efficacy of a novel bipolar radiofrequency ablation device on the beating heart for atrial fibrillation ablation: a long-term porcine study.

作者信息

Voeller Rochus K, Zierer Andreas, Lall Shelly C, Sakamoto Shun-ichiro, Schuessler Richard B, Damiano Ralph J

机构信息

Division of Cardiothoracic Surgery, Washington University School of Medicine, Barnes-Jewish Hospital, St Louis, MO 63110, USA.

出版信息

J Thorac Cardiovasc Surg. 2010 Jul;140(1):203-8. doi: 10.1016/j.jtcvs.2009.06.034. Epub 2010 Feb 1.

Abstract

OBJECTIVE

Over recent years, a variety of energy sources have been used to replace the traditional incisions of the Cox maze procedure for the surgical treatment of atrial fibrillation. This study evaluated the safety and efficacy of a new bipolar radiofrequency ablation device for atrial ablation in a long-term porcine model.

METHODS

Six pigs underwent a Cox maze IV procedure on a beating heart off cardiopulmonary bypass using the AtriCure Isolator II bipolar ablation device (AtriCure, Inc, Cincinnati, Ohio). In addition, 6 pigs underwent median sternotomy and pericardiotomy alone to serve as a control group. All animals were allowed to survive for 30 days. Each pig underwent induction of atrial fibrillation and was then humanely killed to remove the heart en bloc for histologic assessment. Magnetic resonance imaging scans were also obtained preoperatively and postoperatively to assess atrial and ventricular function, pulmonary vein anatomy, valve function, and coronary artery patency.

RESULTS

All animals survived the operation. Electrical isolation of the left atrial appendage and the pulmonary veins was documented by pacing acutely and at 30 days in all animals. No animal that underwent the Cox maze IV procedure was able to be induced into atrial fibrillation at 30 days postoperatively, compared with all the sham animals. All 257 ablations examined were discrete, linear, and transmural, with a mean lesion width of 2.2 +/- 1.1 mm and a mean lesion depth of 5.3 +/- 3.0 mm.

CONCLUSIONS

The AtriCure Isolator II device was able to create reliable long-term transmural lesions of the modified Cox maze procedure on a beating heart without cardiopulmonary bypass 100% of the time. There were no discernible effects on ventricular or valvular function.

摘要

目的

近年来,多种能源已被用于替代传统的Cox迷宫手术切口来治疗心房颤动。本研究在长期猪模型中评估了一种新型双极射频消融装置用于心房消融的安全性和有效性。

方法

6头猪在非体外循环下跳动心脏上使用AtriCure Isolator II双极消融装置(AtriCure公司,俄亥俄州辛辛那提)进行Cox迷宫IV手术。此外,6头猪仅接受正中胸骨切开术和心包切开术作为对照组。所有动物均存活30天。每头猪均诱发心房颤动,然后实施安乐死,完整取出心脏进行组织学评估。术前和术后还进行了磁共振成像扫描,以评估心房和心室功能、肺静脉解剖结构、瓣膜功能和冠状动脉通畅情况。

结果

所有动物均存活至手术结束。所有动物在急性起搏时以及术后30天时均记录到左心耳和肺静脉的电隔离。与所有假手术动物相比,接受Cox迷宫IV手术的动物在术后30天时均不能诱发心房颤动。所检查的257次消融均为离散、线性且透壁的,平均病变宽度为2.2±1.1毫米,平均病变深度为5.3±3.0毫米。

结论

AtriCure Isolator II装置能够在非体外循环下跳动心脏上100%可靠地产生改良Cox迷宫手术的长期透壁病变。对心室或瓣膜功能没有明显影响。

相似文献

引用本文的文献

2
New Biparietal Bipolar Catheter Prototype for Hybrid Atrial Fibrillation Ablation.新型双顶双极导管用于杂交心房颤动消融术。
Innovations (Phila). 2021 Mar-Apr;16(2):181-187. doi: 10.1177/1556984520981025. Epub 2021 Jan 7.
5
Electrophysiological Perspectives on Hybrid Ablation of Atrial Fibrillation.心房颤动杂交消融的电生理视角
J Atr Fibrillation. 2015 Dec 31;8(4):1290. doi: 10.4022/jafib.1290. eCollection 2015 Dec.
9
The impact of previous catheter-based ablation on the efficacy of the Cox-maze IV procedure.既往导管消融对 Cox-maze IV 手术疗效的影响。
Ann Thorac Surg. 2013 Sep;96(3):786-91; discussion 791. doi: 10.1016/j.athoracsur.2013.04.071. Epub 2013 Aug 2.

本文引用的文献

5
Alternative energy sources for the ablation of arrhythmias.
Pacing Clin Electrophysiol. 2005 May;28(5):434-43. doi: 10.1111/j.1540-8159.2005.09481.x.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验