Are home ventilators able to guarantee a minimal tidal volume?

机构信息

Pediatric Pulmonary Department, Research Unit, National Reference Center for Rare Lung Diseases, INSERM, UMR S-938, AP-HP, Hôpital Armand-Trousseau, Université Pierre et Marie Curie-Paris 6, 75012, Paris, France.

出版信息

Intensive Care Med. 2010 Jun;36(6):1008-14. doi: 10.1007/s00134-010-1785-9. Epub 2010 Mar 9.

Abstract

OBJECTIVE

The aim of the study was to evaluate the ability of home ventilators to maintain a minimal tidal volume during different conditions associated with alveolar hypoventilation.

DESIGN

Bench evaluation coupled with an in vivo study on two healthy subjects.

SETTING

Tertiary university hospital and research unit.

INTERVENTIONS

Six ventilators having a "volume guarantee" module (Synchrony II, Philips Respironics; Vivo 40, Breas; Legendair, Airox; Elisee 150, Philips Respironics; Ventimotion, Weinmann; and VS III, Resmed) were tested on a lung bench in a baseline condition and in three conditions associated with alveolar hypoventilation: increase in airway resistance, decrease in lung compliance, and non-intentional leaks. An in vivo study completed the bench study for the non-intentional leak condition.

MEASUREMENTS AND RESULTS

The six ventilators were able to maintain a minimal tidal volume during an increase in airway resistance and a decrease in lung compliance. The maintenance of a minimal tidal volume during a non-intentional leak was more difficult and was associated with large variations in tidal volume, a default of pressure support delivery for some devices, and patient-ventilator dyssynchrony, both during the bench and the in vivo study.

CONCLUSIONS

The six home ventilators tested in the study were able to maintain a minimal tidal volume during an increase in airway resistance and a decrease in lung compliance, but not during a non-intentional leak.

摘要

目的

本研究旨在评估家用呼吸机在与肺泡通气不足相关的不同条件下维持最小潮气量的能力。

设计

结合对两名健康受试者的体内研究进行的台架评估。

地点

三级大学医院和研究单位。

干预措施

对具有“容量保证”模块的六台呼吸机(Synchrony II,飞利浦伟康;Vivo 40,Breas;Legendair,Airox;Elisee 150,飞利浦伟康;Ventimotion, Weinmann;和 VS III,瑞思迈)在肺台上进行测试,分别在基线条件和三种与肺泡通气不足相关的条件下进行测试:气道阻力增加、肺顺应性降低和非故意漏气。体内研究完成了非故意漏气条件下的台架研究。

测量和结果

在气道阻力增加和肺顺应性降低的情况下,六台呼吸机均能维持最小潮气量。在非故意漏气的情况下,维持最小潮气量更加困难,并且与潮气量的较大变化相关,一些设备的压力支持输送出现默认,以及在台架和体内研究期间患者-呼吸机不同步。

结论

在本研究中测试的六台家用呼吸机在气道阻力增加和肺顺应性降低的情况下能够维持最小潮气量,但在非故意漏气的情况下则不能。