J Cataract Refract Surg. 2010 Oct;36(10):1650-5. doi: 10.1016/j.jcrs.2010.05.013.
To compare anterior segment parameters between 1-piece and 3-piece acrylic foldable intraocular lenses (IOLs).
Farabi Eye Hospital, Tehran, Iran.
Prospective randomized comparative case series.
Eyes scheduled for phacoemulsification were randomized into 2 equal groups to receive a 1-piece (AcrySof SA60AT) or 3-piece (AcrySof MA60AC) foldable acrylic IOL. Scheimpflug imaging (Pentacam) was used to measure anterior segment parameters, including anterior chamber depth (ACD), anterior chamber angle (ACA), and anterior chamber volume (ACV), preoperatively and 1 week and 3 months postoperatively.
The study evaluated 125 eyes (123 patients). There was a statistically significant postoperative increase in ACD, ACA, and ACV at 1 week and 3 months in both IOL groups. Although the mean ACD was significantly higher in the 3-piece group at 1 week, there was no significant between-group difference in ACD, ACA, or ACV at 3 months. There was no significant change in anterior segment parameters from 1 week to 3 months in the 1-piece group; however, the 3-piece group had statistically significant decreases in ACD, ACA, and ACV. Refraction remained stable throughout the follow-up in the 1-piece group but showed a significant myopic shift from 1 week to 3 months in the 3-piece group.
After phacoemulsification, the 1-piece acrylic foldable IOL showed little axial movement and provided stable refraction throughout the follow-up. The 3-piece IOL had significant forward movement and led to a myopic shift within 3 months postoperatively. Results indicate that spectacles can be prescribed earlier in eyes with a 1-piece IOL.
比较一体式和三件式丙烯酸可折叠人工晶状体(IOL)的眼前节参数。
伊朗德黑兰法里比眼科医院。
前瞻性随机对照病例系列。
将计划行超声乳化白内障吸除术的患者随机分为两组,分别植入一体式(AcrySof SA60AT)或三件式(AcrySof MA60AC)丙烯酸可折叠人工晶状体。术前及术后 1 周和 3 个月使用 Scheimpflug 成像(Pentacam)测量眼前节参数,包括前房深度(ACD)、前房角(ACA)和前房容积(ACV)。
本研究共纳入 125 只眼(123 例患者)。两组患者术后 1 周和 3 个月时 ACD、ACA 和 ACV 均显著增加。虽然术后 1 周时三组的平均 ACD 显著较高,但 3 个月时两组间 ACD、ACA 或 ACV 无显著差异。一体式组术后 1 周至 3 个月期间眼前节参数无明显变化;而三组患者 ACD、ACA 和 ACV 均呈统计学显著下降。一体式组随访期间屈光度稳定,但三组患者术后 1 周至 3 个月时出现显著近视漂移。
超声乳化白内障吸除术后,一体式丙烯酸可折叠人工晶状体轴向移动较小,随访期间屈光度稳定。三件式人工晶状体有明显的向前移动,术后 3 个月内出现近视漂移。结果表明,植入一体式 IOL 的患者可以更早地配眼镜。