Marini Marino, Da Pozzo Stefano, Accardo Agostino, Canziani Tiziana
Associated Ophthalmic Clinic, Trieste, Italy.
Eur J Ophthalmol. 2011 May-Jun;21(3):258-63. doi: 10.5301/EJO.2010.5767.
To describe Goldmann applanation tonometer (GAT) and rebound tonometer (RT) agreement in measuring intraocular pressure (IOP) in glaucomatous and ocular hypertensive (OH) eyes and to evaluate central corneal thickness (CCT) influence on RT readings.
A total of 347 eyes were enrolled and IOP measured between 9 and 11 am. Rebound tonometry was performed first (RT1), followed by 3 consecutive GAT measurements and by a second RT reading (RT2), within a 5-minute span. Mean IOP (±SD) values were compared by means of paired t-test. Agreement between GAT and RT1 (test 1) and RT2 (test 2) was evaluated with Bland-Altman method, whereas a linear function described the relationship between CCT and IOP taken with RT.
Mean IOP (±SD) taken with RT1, GAT, and RT2 was 18.1±4.3, 15.6±3.3, and 16.3±3.9 mmHg, respectively. Readings were within ±3 mmHg in 63.7% and 86.7% of eyes for test 1 and 2, respectively. A significant (p<0.001) proportional bias was noted on both tests (95% limits of agreement: -2.3/7.4 and -3.6/5.0 mmHg for test 1 and 2, respectively). Agreement between instruments decreased for increasing IOP. Rebound tonometry readings increased by 4.6 and 4.1 mmHg for RT1 and RT2, respectively, for each 100-µm CCT increase.
When used first, RT significantly overestimated IOP compared with GAT. Differences became clinically negligible when RT was used immediately after GAT. Repeated applanation tonometry may explain this observation. Agreement between instruments was acceptable for low IOP, but worsened with increasing IOP values. RT is significantly influenced by CCT. Goldmann applanation tonometer and RT should not be used interchangeably.
描述Goldmann压平眼压计(GAT)和回弹眼压计(RT)在测量青光眼和高眼压症(OH)患者眼压时的一致性,并评估中央角膜厚度(CCT)对RT读数的影响。
共纳入347只眼,于上午9点至11点测量眼压。先进行回弹眼压测量(RT1),随后在5分钟内连续进行3次GAT测量,并进行第二次RT读数(RT2)。采用配对t检验比较平均眼压(±标准差)值。用Bland-Altman方法评估GAT与RT1(测试1)和RT2(测试2)之间的一致性,而用线性函数描述CCT与RT测量的眼压之间的关系。
RT1、GAT和RT2测量的平均眼压(±标准差)分别为18.1±4.3、15.6±3.3和16.3±3.9 mmHg。测试1和测试2中,分别有63.7%和86.7%的眼读数在±3 mmHg范围内。两项测试均发现显著(p<0.001)的比例偏差(测试1和测试2的95%一致性界限分别为-2.3/7.4和-3.6/5.0 mmHg)。随着眼压升高,两种仪器之间的一致性降低。CCT每增加100 µm,RT1和RT2的回弹眼压测量读数分别增加4.6和4.1 mmHg。
与GAT相比,先使用RT时眼压显著高估。在GAT之后立即使用RT时,差异在临床上可忽略不计。多次压平眼压测量可能解释了这一观察结果。两种仪器在低眼压时一致性尚可,但随着眼压值升高而变差。RT受CCT显著影响。Goldmann压平眼压计和RT不应互换使用。