To quantify the amount of aerosol deposited in different parts of the airways with a commercially available nasal sonic jet nebulizer (NJN) using a sound effect, and to compare its performance with a new nasal mesh nebulizer (NMN).
Seven healthy non-smoking male volunteers aged 21-36 years with a mean weight of 77±10 kg were included in this single-center study. Both nebulizer systems were loaded with (99m)Tc-DTPA and scintigraphies were performed with a gamma camera. Particle size distribution of the aerosols produced by the two nebulizer systems was measured.
There was no statistical difference between the two nebulizers in terms of fraction of particles smaller than 5 μm (44±4% vs 45±2%) (p>0.9). Aerosol deposition in the nasal region was 73±10% (% of aerosol deposited in airways) with the NJN, and 99±3% with the NMN (p=0.01). Total nasal deposition was 9.6±1.9% of the nebulizer charge with the NJN and 28.4±8.9% with the NMN (p=0.01). 0.5±0.3% of the nebulizer charge was deposited in the maxillary sinuses with the NJN, compared to 2.2±1.6% with the NMN (p=0.01).
Although the two nebulizers had the same particle size, NMN significantly improved aerosol deposition in nasal cavity and prevents deposition into the lungs.
使用音效定量评估市售鼻用声波喷射雾化器(NJN)和新型鼻用网孔雾化器(NMN)在气道不同部位沉积的气溶胶量,并比较两者的性能。
本单中心研究纳入了 7 名年龄在 21-36 岁之间、平均体重为 77±10kg 的健康非吸烟男性志愿者。这两种雾化器系统均装载(99m)Tc-DTPA 并使用伽马相机进行闪烁显像。测量两种雾化器系统产生的气溶胶的粒径分布。
两种雾化器在小于 5μm 的颗粒比例(44±4% 比 45±2%)方面无统计学差异(p>0.9)。NJN 时,气溶胶在鼻腔中的沉积率为 73±10%(气道沉积的气溶胶百分比),NMN 时为 99±3%(p=0.01)。NJN 的总鼻腔沉积率为雾化器装药量的 9.6±1.9%,NMN 时为 28.4±8.9%(p=0.01)。NJN 时,雾化器装药量的 0.5±0.3%沉积在上颌窦,而 NMN 时为 2.2±1.6%(p=0.01)。
尽管两种雾化器的颗粒大小相同,但 NMN 可显著改善鼻腔内的气溶胶沉积,防止其进入肺部。