Department of Medical Physics, University of Wisconsin, Madison, WI, USA.
Ultrasound Med Biol. 2011 Aug;37(8):1327-39. doi: 10.1016/j.ultrasmedbio.2011.05.023.
Quantitative ultrasound (QUS) has been used to characterize soft tissues at ordinary abdominal ultrasound frequencies (2 to 15 MHz) and is beginning application at high frequencies (20 to 70 MHz). For example, backscatter and attenuation coefficients can be estimated in vivo using a reference phantom. At high frequencies, it is crucial that reverberations do not compromise the measurements. Such reverberations can occur between the phantom's scanning window and transducer components as well as within the scanning window between its surfaces. Transducers are designed to minimize reverberations between the transducer and soft tissue. Thus, the acoustic impedance of a phantom scanning window should be tissuelike; polymethylpentene (TPX) is commonly used because of its tissuelike acoustic impedance. For QUS, it is also crucial to correct for the transmission coefficient of the scanning window. Computation of the latter requires knowledge of the ultrasonic properties, viz, density, speed and attenuation coefficients. This work reports values for the ultrasonic properties of two versions of TPX over the high-frequency range. One form (TPX film) is used as a scanning window on high-frequency phantoms, and at 40 MHz and 22°C was found to have an attenuation coefficient of 120 dB/cm and a propagation speed of 2093 m/s.
定量超声(QUS)已被用于描述普通腹部超声频率(2 至 15MHz)下的软组织,并开始在高频(20 至 70MHz)下应用。例如,可以使用参考体模在体内估算背散射和衰减系数。在高频下,至关重要的是避免混响影响测量。这种混响可能发生在体模的扫描窗口和换能器组件之间,以及扫描窗口的表面之间。换能器的设计旨在最小化换能器和软组织之间的混响。因此,体模扫描窗口的声阻抗应该类似于组织;聚甲基戊烯(TPX)因其类似于组织的声阻抗而被广泛使用。对于 QUS,校正扫描窗口的传输系数也很关键。后者的计算需要了解超声波特性,即密度、速度和衰减系数。这项工作报告了两种版本的 TPX 在高频范围内的超声波特性值。其中一种形式(TPX 薄膜)用作高频体模的扫描窗口,在 40MHz 和 22°C 时,衰减系数为 120dB/cm,传播速度为 2093m/s。