Make-up of the epinephrine autoinjector: the effect on its use by untrained users.

机构信息

Department of Pediatric Allergy and Asthma, Gazi University Faculty of Medicine, Besevler, Ankara, Turkey.

出版信息

Pediatr Allergy Immunol. 2011 Nov;22(7):729-33. doi: 10.1111/j.1399-3038.2011.01195.x. Epub 2011 Jul 13.

Abstract

BACKGROUND

Part of the problems related to proper use of the epinephrine autoinjector may be related to the design of the autoinjector itself.

AIM

We investigated whether minor modifications in the design of the currently available epinephrine autoinjector ease its use and abrogate common use errors.

METHODS

All interns other than those who had previously worked in allergy department in a medical school were invited to the study. Two identical epinephrine autoinjector trainers (Epipen trainer(®) ) were used, one of which was modified by changing the gray safety cap to red and placing a yellow arrow pointing to the black injection tip. A written and visual instruction sheet for each trainer was provided. Participants were asked to demonstrate the use of the Epipen trainer either with the original or with the modified one. They were scored and timed for their demonstration.

RESULTS

Out of the 224 interns who were invited to participate, one hundred and sixty-four interns (73.2%) participated in the study. The number of participants correctly demonstrating the use of epinephrine autoinjectors was 22.6% and 65% in unmodified and modified trainer groups, respectively (p < 0.001). The mean time to administer trainers was 26.78 ± 10.6 and 15.88 ± 2.55 s; total median scores were 3.08 ± 1.48 and 4.47 ± 0.84 in unmodified and modified groups, respectively (p < 0.001 for both). Significantly fewer participants had presumptive unintentional injection injury while using modified (5%) compared with unmodified trainer (45.2%) (p < 0.001).

CONCLUSION

Few and simple modifications in the design of epinephrine autoinjector were found effective in increasing its correct use and decreasing common use errors by untrained users. (Clinical trials identifier: NCT01217138).

摘要

背景

适当使用肾上腺素自动注射器相关问题的一部分可能与自动注射器本身的设计有关。

目的

我们旨在研究肾上腺素自动注射器设计方面的微小改变是否可以使其使用更简便,并消除常见的使用错误。

方法

除了在医学院过敏科工作过的实习医生外,我们邀请了所有的实习医生参加这项研究。使用了两个相同的肾上腺素自动注射器训练器(EpiPen 训练器),其中一个通过将灰色安全帽改为红色并在黑色注射尖端处贴上黄色箭头来进行修改。为每个训练器都提供了书面和视觉使用说明。要求参与者用原始训练器或修改后的训练器演示 EpiPen 训练器的使用。对他们的演示进行评分和计时。

结果

在受邀参加的 224 名实习医生中,有 164 名(73.2%)参加了研究。正确演示肾上腺素自动注射器使用方法的参与者人数分别为未修改组和修改组的 22.6%和 65%(p < 0.001)。使用训练器的平均时间分别为 26.78 ± 10.6 和 15.88 ± 2.55 秒;总中位数评分分别为未修改组和修改组的 3.08 ± 1.48 和 4.47 ± 0.84(均 p < 0.001)。在使用修改后的训练器时,只有 5%的参与者有假定的非故意注射损伤,而使用未修改的训练器时,这一比例为 45.2%(p < 0.001)。

结论

在肾上腺素自动注射器的设计方面进行了少量且简单的修改,发现这可以有效提高未经培训的使用者的正确使用率,并减少常见的使用错误。(临床试验标识符:NCT01217138)。