眼前节分析系统 Orbscan II、Pentacam 和 Galilei 角膜地形图仪测量的角膜曲率和角膜屈光度的比较及可重复性。

Comparison and repeatability of keratometric and corneal power measurements obtained by Orbscan II, Pentacam, and Galilei corneal tomography systems.

机构信息

Department of Ophthalmology, New Zealand National Eye Centre, Faculty of Medical and Health Sciences, University of Auckland, Auckland, New Zealand.

出版信息

Am J Ophthalmol. 2013 Jul;156(1):53-60. doi: 10.1016/j.ajo.2013.01.029. Epub 2013 Mar 28.

Abstract

PURPOSE

To assess the repeatability and comparability of corneal power and central corneal thickness (CCT) measurements obtained using Orbscan II (Bausch & Lomb), Pentacam (Oculus), and Galilei (Ziemer) tomographers.

DESIGN

Prospective, comparative study.

METHODS

setting: Departments of Ophthalmology, University of Auckland and Auckland District Health Board, Auckland, New Zealand. study population: Thirty eyes of 30 healthy participants. observations. CCT and corneal power measured using Orbscan II, Pentacam, and Galilei tomography. main outcome measures: Degree of agreement in and repeatability of CCT and corneal power measures.

RESULTS

Orbscan II measured significantly lower CCT compared with Pentacam (20 μm; P < .0005) and Galilei (18 μm; P < .0005). The Orbscan II had wide limits of agreement when compared with both the Galilei (-11 to 47 μm) and Pentacam (-88 to 47 μm). For each device, the intraclass correlation coefficient for CCT was higher than 0.9. The coefficient of variation ranged from 0.33% to 0.93%. There was no significant difference in mean steep keratometry or mean flat keratometry between instrument pairs. However, there was poor agreement in flat keratometry and steep keratometry obtained by Orbscan II compared with those obtained by the Galilei and Pentacam.

CONCLUSIONS

The keratometry and pachymetry measurements obtained by Orbscan II, Pentcam, and Galilei tomographers were sufficiently disparate that the 3 devices could not be considered equivalent. All 3 devices demonstrated a high level of repeatability, although the Galilei exhibited the best repeatability.

摘要

目的

评估 Orbscan II(博士伦)、Pentacam(欧科路)和 Galilei(齐默)角膜地形图仪测量角膜屈光力和中央角膜厚度(CCT)的可重复性和可比性。

设计

前瞻性、对比研究。

方法

地点:新西兰奥克兰大学和奥克兰地区卫生委员会眼科系。研究人群:30 名健康参与者的 30 只眼。观察指标:使用 Orbscan II、Pentacam 和 Galilei 断层扫描仪测量 CCT 和角膜屈光力。主要观察指标:CCT 和角膜屈光力测量的一致性和可重复性。

结果

Orbscan II 测量的 CCT 明显低于 Pentacam(20μm;P<0.0005)和 Galilei(18μm;P<0.0005)。与 Galilei(-11 至 47μm)和 Pentacam(-88 至 47μm)相比,Orbscan II 的一致性范围很宽。对于每种设备,CCT 的组内相关系数均高于 0.9。变异系数范围为 0.33%至 0.93%。仪器对之间平均陡峭角膜曲率和平均平坦角膜曲率无显著差异。然而,与 Galilei 和 Pentacam 相比,Orbscan II 获得的平坦角膜曲率和陡峭角膜曲率一致性较差。

结论

Orbscan II、Pentacam 和 Galilei 角膜地形图仪测量的角膜屈光力和角膜厚度差异很大,因此这 3 种设备不能视为等效。所有 3 种设备均表现出高度的可重复性,尽管 Galilei 的可重复性最好。