This study was conducted to evaluate the level of agreement in keratometry measurements between a rotating Scheimpflug imaging-based system (Pentacam) and a handheld auto-refractokeratometer (handheld NIDEK ARK-30).
This analytical cross-sectional study was conducted in the right eyes of 579 subjects. Keratometry measurements were conducted with the Pentacam and the handheld NIDEK ARK-30 systems. The SPSS Software version 22 and MedCalc V3 were applied to estimate descriptive statistics using paired t-test, Pearson correlation coefficient, 95% limits of agreement (LoA), and Bland-Altman plot.
In the total sample, the inter-device difference in the mean flat and steep keratometry values was -0.266 diopter (D) (P-value<0.001) and 0.052D (P-value=0.093), respectively. There was a significant difference in mean flat keratometry between the two devices in all groups of refractive errors (paired difference <0.5D and P-value<0.001). The difference in mean steep keratometry was significant only in myopic subjects (P-value=0.046). The 95% LoA between the two devices measurements was 2.51D, 3.98D, and 6.37D for flat keratometry and 2.6D, 3.2D, and 3.9D for steep keratometry in emmetropic, myopic, and hyperopic subjects, respectively.
Our study showed relatively wide limits of agreement between handheld NIDEK ARK-30 and Pentacam; therefore, these devices cannot be used interchangeably for measuring corneal curvature.
本研究旨在评估基于旋转Scheimpflug成像系统(Pentacam)和手持自动验光角膜曲率计(手持NIDEK ARK - 30)之间角膜曲率测量的一致性水平。
本分析性横断面研究在579名受试者的右眼进行。使用Pentacam和手持NIDEK ARK - 30系统进行角膜曲率测量。应用SPSS软件22版和MedCalc V3通过配对t检验、Pearson相关系数、95%一致性界限(LoA)和Bland - Altman图来估计描述性统计量。
在整个样本中,平均平坦和陡峭角膜曲率值的设备间差异分别为 - 0.266屈光度(D)(P值<0.001)和0.052D(P值 = 0.093)。在所有屈光不正组中,两种设备之间的平均平坦角膜曲率存在显著差异(配对差异<0.5D且P值<0.001)。平均陡峭角膜曲率的差异仅在近视受试者中显著(P值 = 0.046)。对于正视、近视和远视受试者,两种设备测量的平坦角膜曲率的95% LoA分别为2.51D、3.98D和6.37D,陡峭角膜曲率的95% LoA分别为2.6D、3.2D和3.9D。
我们的研究表明手持NIDEK ARK - 30和Pentacam之间的一致性界限相对较宽;因此,这些设备不能互换用于测量角膜曲率。