Evaluating and modeling of photon beam attenuation by a standard treatment couch.

机构信息

Department of Radiation Oncology, Cancer Institute & Hospital, Chinese Academy of Medical Sciences, Beijing, China.

出版信息

J Appl Clin Med Phys. 2011 Jul 12;12(4):3561. doi: 10.1120/jacmp.v12i4.3561.

Abstract

The purpose of this study was to evaluate beam attenuation by treatment couch and build a treatment couch model in TPS to check for beam-couch intersection at the planning stage and deal with beam attenuation by treatment couch in dose calculation. In this study, a standard treatment couch, Siemens ZXT couch, has been incorporated into Pinnacle3 8.0 TPS, based on an existing TPS tool, model-based segmentation (MBS). This was done by generating the couch's model from contours of the couch, together with the density information. Both the geometric and dosimetric accuracy of the couch model were evaluated. The test of beam-couch intersection prediction showed good agreement between predicted and measured results, and the differences were within 1° gantry rotation. For individual posterior oblique beams, the attenuation by metallic frames and PMMA couch top could reach nearly as high as 60% and 10%, respectively. For several posterior oblique beams (180°, 220°, 235°) that attenuated by the PMMA couch top, the calculated and measured dose distributions were compared. The dose differences at central axis were within 1%, and almost all points agreed with the calculations when the DD and DTA criteria of 3%/3 mm were adopted. The difference between calculated and measured attenuation factors were within 0.5%. This study demonstrates that the couch model created by MBS, which contains geometric and density information of the couch, can be used to detect the beam-couch intersection, and also is able to provide an accurate representation of the couch top attenuation properties in patient dose calculation.

摘要

本研究旨在评估治疗床对射束的衰减,并在治疗计划系统 (TPS) 中建立治疗床模型,以在规划阶段检查射束-治疗床的交叉,并在剂量计算中处理治疗床的射束衰减。在本研究中,基于现有的基于模型的分割(MBS)TPS 工具,将标准治疗床(西门子 ZXT 治疗床)整合到 Pinnacle3 8.0 TPS 中。这是通过从治疗床的轮廓以及密度信息生成治疗床的模型来实现的。评估了治疗床模型的几何和剂量学准确性。对射束-治疗床交叉预测的测试表明,预测结果与测量结果之间具有良好的一致性,差异在 1°旋转范围内。对于个别后斜射束,金属框架和 PMMA 治疗床顶部的衰减率可分别达到近 60%和 10%。对于几个由 PMMA 治疗床顶部衰减的后斜射束(180°、220°、235°),比较了计算和测量的剂量分布。在中心轴上的剂量差异在 1%以内,当采用 3%/3mm 的 DD 和 DTA 标准时,几乎所有点都与计算结果一致。计算和测量衰减因子之间的差异在 0.5%以内。本研究表明,由 MBS 创建的包含治疗床几何和密度信息的治疗床模型可用于检测射束-治疗床的交叉,并且还能够在患者剂量计算中准确表示治疗床顶部的衰减特性。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/6547/5718748/93e899d14ee9/ACM2-12-139-g001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验