J Clin Rheumatol. 2012 Jan;18(1):23-7. doi: 10.1097/RHU.0b013e31823e5cda.
Dual-energy computed tomography (DECT) has recently been described as a sensitive method to detect urate deposits in patients with gout.
The aim of this study was to compare the reproducibility of DECT with various physical measurement methods of tophus size assessment.
Sixty-four tophi from 25 patients were analyzed. Each tophus was assessed by 2 independent observers using Vernier calipers and tape measure. All patients proceeded to DECT scanning of both feet. Urate volume within index tophi was assessed by 2 independent observers using automated DECT volume assessment software (n = 55 tophi). Five patients returned within 1 week for repeat physical assessment of tophus size. Dual-energy computed tomography scans from the returning patients were scored twice by both observers. Intraobserver and interobserver reproducibility was assessed by intraclass correlation coefficient (ICC) and limits-of-agreement analysis.
Overall, DECT was more reproducible than the physical methods with interobserver ICCs for DECT of 0.95, for calipers 0.78, and for tape measurement 0.88, and intraobserver ICCs for DECT of 1.00, for calipers 0.75, and for tape measurement 0.91. Vernier caliper and tape measurements correlated highly with each other (rs = 0.84, P < 0.0001) but less well with DECT (for index tophi, r(s) = 0.46, P = 0.004 for both). Large variation was observed in the amount of urate deposits documented by DECT in tophi of similar physical size.
Dual-energy computed tomography scanning is a highly reproducible method for measuring urate deposits within tophi. This imaging modality reveals the composition of tophi that contain variable urate deposits embedded within soft tissue.
双能 CT(DECT)最近被描述为一种检测痛风患者尿酸盐沉积的敏感方法。
本研究旨在比较 DECT 与各种尿酸盐沉积大小评估物理测量方法的重现性。
对 25 例患者的 64 个痛风石进行分析。每个痛风石均由 2 名独立观察者使用游标卡尺和卷尺进行评估。所有患者均进行双脚 DECT 扫描。由 2 名独立观察者使用自动 DECT 容积评估软件评估尿酸盐容积(n = 55 个痛风石)。5 例患者在 1 周内再次进行痛风石大小的重复体格评估。返回的患者的 DECT 扫描由两位观察者两次评分。采用组内相关系数(ICC)和界限协议分析评估观察者内和观察者间的可重复性。
总体而言,DECT 的观察者间 ICC 为 0.95,游标卡尺为 0.78,卷尺为 0.88,均优于物理方法,观察者内 ICC 为 1.00,游标卡尺为 0.75,卷尺为 0.91。游标卡尺和卷尺之间的相关性很高(rs = 0.84,P <0.0001),但与 DECT 的相关性较差(对于指数痛风石,r(s) = 0.46,P = 0.004)。在物理大小相似的痛风石中,DECT 记录的尿酸盐沉积量存在较大差异。
DECT 扫描是一种测量痛风石内尿酸盐沉积的高度可重复的方法。这种成像方式揭示了含有不同尿酸盐沉积的软组织内痛风石的组成。