Suppr超能文献

基于迭代运动校正的实时 MRI 高时间分辨率电影图像的回顾性重建。

Retrospective reconstruction of high temporal resolution cine images from real-time MRI using iterative motion correction.

机构信息

National Heart, Lung, and Blood Institute, National Institutes of Health, Bethesda, Maryland, USA.

出版信息

Magn Reson Med. 2012 Sep;68(3):741-50. doi: 10.1002/mrm.23284. Epub 2011 Dec 21.

Abstract

Cardiac function has traditionally been evaluated using breath-hold cine acquisitions. However, there is a great need for free breathing techniques in patients who have difficulty in holding their breath. Real-time cardiac MRI is a valuable alternative to the traditional breath-hold imaging approach, but the real-time images are often inferior in spatial and temporal resolution. This article presents a general method for reconstruction of high spatial and temporal resolution cine images from a real-time acquisition acquired over multiple cardiac cycles. The method combines parallel imaging and motion correction based on nonrigid registration and can be applied to arbitrary k-space trajectories. The method is demonstrated with real-time Cartesian imaging and Golden Angle radial acquisitions, and the motion-corrected acquisitions are compared with raw real-time images and breath-hold cine acquisitions in 10 (N = 10) subjects. Acceptable image quality was obtained in all motion-corrected reconstructions, and the resulting mean image quality score was (a) Cartesian real-time: 2.48, (b) Golden Angle real-time: 1.90 (1.00-2.50), (c) Cartesian motion correction: 3.92, (d) Radial motion correction: 4.58, and (e) Breath-hold cine: 5.00. The proposed method provides a flexible way to obtain high-quality, high-resolution cine images in patients with difficulty holding their breath.

摘要

心脏功能传统上采用屏气电影采集进行评估。然而,对于那些难以屏气的患者,非常需要自由呼吸技术。实时心脏 MRI 是传统屏气成像方法的一种有价值的替代方法,但实时图像在空间和时间分辨率上往往较差。本文提出了一种从多个心脏周期采集的实时采集重建高空间和高时间分辨率电影图像的通用方法。该方法结合了基于非刚性配准的并行成像和运动校正,可应用于任意 k 空间轨迹。该方法通过实时笛卡尔成像和 Golden Angle 径向采集进行了演示,并将运动校正采集与原始实时图像和屏气电影采集在 10 名(N = 10)受试者中进行了比较。在所有运动校正重建中都获得了可接受的图像质量,并且得到的平均图像质量评分分别为(a)笛卡尔实时:2.48;(b)Golden Angle 实时:1.90(1.00-2.50);(c)笛卡尔运动校正:3.92;(d)径向运动校正:4.58;以及(e)屏气电影采集:5.00。该方法为在难以屏气的患者中获得高质量、高分辨率电影图像提供了一种灵活的方法。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/96b2/3311753/91b19506af78/nihms333085f1.jpg

相似文献

引用本文的文献

9
Cardiac MR: From Theory to Practice.心脏磁共振成像:从理论到实践。
Front Cardiovasc Med. 2022 Mar 3;9:826283. doi: 10.3389/fcvm.2022.826283. eCollection 2022.

本文引用的文献

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验