Three-dimensional surface scanning for accurate patient positioning and monitoring during breast cancer radiotherapy.

机构信息

Department of Radiotherapy and Radio-Oncology, Landeskrankenhaus Salzburg, Paracelsus Medical University Clinics, Müllner Hauptstr. 48, 5020, Salzburg, Austria,

出版信息

Strahlenther Onkol. 2013 Oct;189(10):887-93. doi: 10.1007/s00066-013-0358-6. Epub 2013 Jun 7.

Abstract

PURPOSE

Clinical evaluation of an optical three-dimensional surface scanning (3D-SS) system for patient positioning and monitoring during radiotherapy (RT) for breast cancer.

MATERIALS AND METHODS

A ceiling-mounted scanner was developed to acquire multiple 3D body surface images and tested in 14 conservatively operated breast cancer patients. A reference skin surface was derived from the planning computed tomography (CT) scan as basis for rigid registration with the surface scans. In addition to electronic portal images (EPIs), optical scans were acquired at three defined time points before and during daily RT. Patient setup was guided by laser alignments and corrected according to EPI findings. The accuracy of the 3D-SS system was validated by comparison of the optical scans to EPIs generated in parallel. Interfraction shifts were investigated by comparison of the first 3D-SS image with the reference body outline. Intrafractional motions were analysed by comparing the three daily surface scans with the first EPI.

RESULTS

Comparison of EPIs and 3D-SS images revealed good accordance (- 0.05±0.94 mm). Analysis of daily patient positions revealed average deviations of 0.4±2.4 mm laterally, 0.3±1.9 mm longitudinally and 0.2±3.3 mm vertically. After 2 weeks, a systematic interfraction shift in patient positioning was noted, particularly in the vertical direction (4.9±0.56 mm), which was attributed to patients progressively relaxing. 3D-SS images showed intrafractional shifts of 1.2±0.7 mm over a time course of 2 min.

CONCLUSION

Optical surface scanning is a simple, fast and reproducible method for breast cancer patient alignment. Particularly for more sophisticated irradiation techniques, it helps to improve accuracy in patient positioning during radiotherapy without the exposure to additional ionizing radiation.

摘要

目的

临床评估一种光学三维表面扫描(3D-SS)系统,用于乳腺癌放射治疗(RT)期间的患者定位和监测。

材料和方法

开发了一种天花板安装的扫描仪,用于获取多个 3D 体表面图像,并在 14 名接受保乳手术的乳腺癌患者中进行了测试。参考皮肤表面是从计划计算机断层扫描(CT)扫描中得出的,作为与表面扫描进行刚性配准的基础。除了电子射野影像(EPI)外,在每天 RT 之前和期间的三个定义时间点采集光学扫描。根据激光对准引导患者设置,并根据 EPI 结果进行校正。通过将光学扫描与平行生成的 EPI 进行比较,验证了 3D-SS 系统的准确性。通过比较第一次 3D-SS 图像与参考身体轮廓,研究了分次间移位。通过比较三次日常表面扫描与第一次 EPI,分析了分次内运动。

结果

EPI 和 3D-SS 图像的比较显示出良好的一致性(-0.05±0.94mm)。对每日患者位置的分析显示,横向平均偏差为 0.4±2.4mm,纵向平均偏差为 0.3±1.9mm,垂直平均偏差为 0.2±3.3mm。2 周后,患者定位出现系统性分次间移位,特别是在垂直方向(4.9±0.56mm),这归因于患者逐渐放松。3D-SS 图像显示在 2 分钟的时间过程中,分次内移位为 1.2±0.7mm。

结论

光学表面扫描是一种简单、快速且可重复的乳腺癌患者对齐方法。对于更复杂的照射技术,它有助于在不增加电离辐射暴露的情况下提高放射治疗期间患者定位的准确性。