Int J Audiol. 2012 Aug;51(8):606-10. doi: 10.3109/14992027.2012.688144. Epub 2012 Jun 7.
To establish the reliability and validity of a computerized self-administered hearing test.
Cross-sectional within a comparative study of subjects.
Subjects were 100 Chinese adults who attended the audiology clinic in a hospital for a hearing test.
There was no significant difference in the thresholds of unmasked air-conduction hearing obtained with the computerized self-administered hearing test via a smartphone and those obtained with standard pure-tone audiometry. High test-retest reliability was observed with the self-administered hearing test (intraclass correlation coefficient = 0.95), and was comparable with that observed in standard pure-tone audiometry (intraclass correlation coefficient = 0.97). The thresholds of the self-administered hearing test measured in a sound-proof booth were not significantly different from those measured in a quiet office room.
The results suggest that the computerized self-administered hearing test is a reliable and valid measure of unmasked air-conduction hearing thresholds.