Ultrasound Med Biol. 2012 Aug;38(8):1476-82. doi: 10.1016/j.ultrasmedbio.2012.04.004. Epub 2012 Jun 12.
The induction of pulmonary hemorrhage by pulsed ultrasound was discovered over 20 years ago. This phenomenon may pose a risk of patient lung injury, particularly for point of care pulmonary ultrasound. A diagnostic ultrasound machine (HDI 5000; Philips Healthcare, Andover MA USA) with 7.6 MHz (CL15-7) linear array was used to image the right lung of anesthetized rats in a warmed water bath. The image showed rapid initiation and progression of comet tail artifacts across the lung image for an on-screen mechanical index (MI) of 0.9, which corresponded to a pulmonary hemorrhage in the lung. Groups of rats were scanned at a range of MI settings and a threshold was located at an MI of about 0.44. This finding indicated a greater sensitivity to pulmonary ultrasound than was expected, based on previous results. Further research is needed to understand this phenomenon and to develop safety guidelines for sonographers.
脉冲超声诱发肺出血是 20 多年前发现的。这种现象可能会对患者的肺部造成损伤,尤其是在即时护理肺部超声中。使用带有 7.6MHz(CL15-7)线阵探头的诊断超声机(HDI 5000;飞利浦医疗,美国安多弗)对温水浴中麻醉大鼠的右肺进行成像。图像显示,在屏幕机械指数(MI)为 0.9 时,彗尾伪影迅速在整个肺图像上起始和进展,这对应于肺中的肺出血。对不同 MI 设置的大鼠组进行了扫描,并在 MI 约为 0.44 处定位了一个阈值。这一发现表明,与以前的结果相比,对肺超声的敏感性更高。需要进一步研究以了解这一现象,并为超声医师制定安全指南。