Suppr超能文献

19 氟磁共振血管造影术在小鼠中的应用。

Fluorine-19 magnetic resonance angiography of the mouse.

机构信息

Department of Radiology, Centre Hospitalier Universitaire Vaudois and University of Lausanne, Lausanne, Switzerland.

出版信息

PLoS One. 2012;7(7):e42236. doi: 10.1371/journal.pone.0042236. Epub 2012 Jul 27.

Abstract

PURPOSE

To implement and characterize a fluorine-19 ((19)F) magnetic resonance imaging (MRI) technique and to test the hypothesis that the (19)F MRI signal in steady state after intravenous injection of a perfluoro-15-crown-5 ether (PCE) emulsion may be exploited for angiography in a pre-clinical in vivo animal study.

MATERIALS AND METHODS

In vitro at 9.4T, the detection limit of the PCE emulsion at a scan time of 10 min/slice was determined, after which the T(1) and T(2) of PCE in venous blood were measured. Permission from the local animal use committee was obtained for all animal experiments. 12 µl/g of PCE emulsion was intravenously injected in 11 mice. Gradient echo (1)H and (19)F images were obtained at identical anatomical levels. Signal-to-noise (SNR) and contrast-to-noise (CNR) ratios were determined for 33 vessels in both the (19)F and (1)H images, which was followed by vessel tracking to determine the vessel conspicuity for both modalities.

RESULTS

In vitro, the detection limit was ∼400 µM, while the (19)F T(1) and T(2) were 1350±40 and 25±2 ms. The (19)F MR angiograms selectively visualized the vasculature (and the liver parenchyma over time) while precisely coregistering with the (1)H images. Due to the lower SNR of (19)F compared to (1)H (17±8 vs. 83±49, p<0.001), the (19)F CNR was also lower at 15±8 vs. 52±35 (p<0.001). Vessel tracking demonstrated a significantly higher vessel sharpness in the (19)F images (66±11 vs. 56±12, p = 0.002).

CONCLUSION

(19)F magnetic resonance angiography of intravenously administered perfluorocarbon emulsions is feasible for a selective and exclusive visualization of the vasculature in vivo.

摘要

目的

实现并描述一种氟-19((19)F)磁共振成像(MRI)技术,并验证静脉注射全氟-15-冠-5 醚(PCE)乳剂后,(19)F MRI 信号在稳态下可用于临床前动物研究中的血管造影的假说。

材料与方法

在 9.4T 下进行体外实验,确定 PCE 乳剂在扫描时间为 10 分钟/切片时的检测限,然后测量静脉血中 PCE 的 T1 和 T2。所有动物实验均获得当地动物使用委员会的许可。11 只小鼠静脉注射 12μl/g PCE 乳剂。在相同的解剖水平获得梯度回波(1)H 和(19)F 图像。在(19)F 和(1)H 图像中,确定了 33 个血管的信噪比(SNR)和对比噪声比(CNR)比值,随后进行血管跟踪以确定两种模态的血管显影效果。

结果

在体外,检测限约为 400μM,而(19)F T1 和 T2 分别为 1350±40ms 和 25±2ms。(19)F MR 血管造影术选择性地可视化了血管(并随着时间推移可视化了肝实质),同时与(1)H 图像精确配准。由于(19)F 的 SNR 低于(1)H(17±8 与 83±49,p<0.001),(19)F 的 CNR 也较低,为 15±8 与 52±35(p<0.001)。血管跟踪显示,(19)F 图像中的血管锐度明显更高(66±11 与 56±12,p=0.002)。

结论

静脉注射全氟碳乳剂的(19)F 磁共振血管造影术可用于选择性和排他性地可视化体内血管。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/935e/3407132/c750ae57ae87/pone.0042236.g001.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验