Kruskall M S, Pacini D G, Malynn E R, Button L N
Department of Pathology, Charles A. Dana Research Institute, Harvard Thorndike Laboratory, Beth Israel Hospital, Boston, Massachusetts.
Transfusion. 1990 Jan;30(1):7-10. doi: 10.1046/j.1537-2995.1990.30190117636.x.
Most commercial blood warmers are unable to warm blood components to greater than 35 degrees C at rapid flow rates (greater than 100 mL/min) because of problems with inefficient heat transfer and resistance to the flow of blood through the instrument's tubing. A new blood warmer, using a closed-flow 40 degrees C counter-current water bath and large-bore tubing, is able to warm cold (10 degrees C) packed red cells to temperatures greater than 35 degrees C even at flow rates of 500 mL per minute. To evaluate the effects on red cells of prolonged exposure to the 40 degrees C heat exchanger, blood from five volunteers stored for 42 days in AS-1 was run through the blood warmer using flow rates equivalent to the transfusion of a unit of blood over 4 hours. Compared with unwarmed blood, these blood units showed no significant changes in plasma hemoglobin, mean corpuscular hemoglobin concentration, potassium, ATP, pH, and osmotic fragility. In vivo survivals of chromium-51-labeled warmed autologous red cells were greater than 75 percent in four volunteers and 49.5 percent in the fifth volunteer, whose abnormally low unwarmed red cell ATP level suggested a storage problem. Thus, this blood warmer, which warms blood efficiently at high flow rates, causes no red cell damage, even with prolonged exposure of old red cells to the 40 degrees C heat exchanger.
大多数商用血液加温器由于存在热传递效率低下以及血液流经仪器管道时存在阻力等问题,无法在快速流速(大于100毫升/分钟)下将血液成分加温至高于35摄氏度。一种新型血液加温器采用40摄氏度的闭流式逆流水浴和大口径管道,即使在流速为每分钟500毫升的情况下,也能将冷的(10摄氏度)浓缩红细胞加温至高于35摄氏度。为了评估长时间暴露于40摄氏度热交换器对红细胞的影响,将5名志愿者在AS - 1中储存42天的血液以相当于4小时输注一个单位血液的流速通过血液加温器。与未加温的血液相比,这些血液单位在血浆血红蛋白、平均红细胞血红蛋白浓度、钾、三磷酸腺苷(ATP)、pH值和渗透脆性方面均无显著变化。在4名志愿者中,经铬 - 51标记的加温自体红细胞的体内存活率大于75%,在第5名志愿者中为49.5%,其未加温红细胞的ATP水平异常低表明存在储存问题。因此,这种能在高流速下有效加温血液的血液加温器,即使长时间让陈旧红细胞暴露于40摄氏度的热交换器中,也不会造成红细胞损伤。