Patel N, Smith C E, Pinchak A C
Department of Anesthesia, MetroHealth Medical Center, Case Western Reserve University, Cleveland, Ohio 44109, USA.
Can J Anaesth. 1995 Jul;42(7):636-42. doi: 10.1007/BF03011886.
The study evaluated the warming ability and flow rates associated with four fluid warming devices during pressure driven infusion and during wide open gravity driven roller clamp infusion. Warmers tested were the Astotherm, Flotem IIe, Level 1 System 250 and a modified cardioplegia heat exchanger. Fluids tested were crystalloid, red cells diluted with 200 ml, 0.9% saline, and undiluted red cells. The volume of fluid and outlet temperatures (point where i.v. tubing would be attached to the patient) were measured for each fluid and compared among warmers for each flow rate condition. For pressure driven infusion of red cells and crystalloid, the System 250, and modified heat exchanger delivered warmer fluids (33-35 degrees C) at higher flow rates (160-740 ml.min-1) than the Astotherm and Flotem (23-31 degrees C, 44-268 ml.min-1, P < 0.05). For gravity driven infusion, the System 250 delivered the warmest fluids (33-36 degrees C, P < 0.05) compared with the modified heat exchanger (29-35 degrees C), Astotherm (26-32 degrees C) and Flotem (26-27 degrees C). In conclusion, the modified heat exchanger and System 250 were moderately effective (outlet temperature > 32 degrees C) in warming crystalloid and red cells at pressure driven flow rates. Only the System 250 warmed red cells > 35 degrees C at gravity driven flow rates. The Flotem and Astotherm were not effective in warming rapidly infused solutions. None of the warmers tested was able to deliver fluids at normothermia (> 36.5 degrees C).
该研究评估了四种输液加温设备在压力驱动输注和大开重力驱动滚轮夹输注过程中的加温能力及流速。所测试的加温器有Astotherm、Flotem IIe、Level 1 System 250和一种改良的心脏停搏液热交换器。所测试的液体有晶体液、用200毫升0.9%生理盐水稀释的红细胞以及未稀释的红细胞。测量了每种液体的液体量和出口温度(静脉输液管连接到患者的部位),并在每种流速条件下对各加温器进行比较。对于压力驱动输注红细胞和晶体液,System 250和改良热交换器在较高流速(160 - 740毫升·分钟-1)下能输送温度更高的液体(33 - 35摄氏度),高于Astotherm和Flotem(23 - 31摄氏度,44 - 268毫升·分钟-1,P < 0.05)。对于重力驱动输注,与改良热交换器(29 - 35摄氏度)、Astotherm(26 - 32摄氏度)和Flotem(26 - 27摄氏度)相比,System 250输送的液体温度最高(33 - 36摄氏度,P < 0.05)。总之,改良热交换器和System 250在压力驱动流速下加温晶体液和红细胞方面有一定效果(出口温度> 32摄氏度)。只有System 250在重力驱动流速下能将红细胞加温至> 35摄氏度。Flotem和Astotherm在快速输注溶液的加温方面效果不佳。所测试的加温器均无法在正常体温(> 36.5摄氏度)下输送液体。