Michaelsen Kelly, Krishnaswamy Venkataramanan, Pogue Brian W, Brooks Ken, Defreitas Ken, Shaw Ian, Poplack Steven P, Paulsen Keith D
Thayer School of Engineering, Dartmouth College, Hanover, NH 03755, USA.
Biomed Opt Express. 2012 Sep 1;3(9):2078-86. doi: 10.1364/BOE.3.002078. Epub 2012 Aug 10.
Development of a detector case for complete co-registration of images in a non-fiber-based combined near-infrared spectral tomography and digital breast tomosynthesis, required analysis to find materials that could support a breast under full mammographic compression without affecting the x-ray images or the quality of the near infrared measurements. Several possible solutions were considered, and many types of plastics were tested in the development of the detector case. Light channeling within the detector case changed the data obtained in resin and agarose phantoms, lowering recovered absorption values. Additional developments focusing on blocking stray light were successful and permitted a normal subject imaging exam.