Oral Radiology Area, Department of Oral Diagnosis, Piracicaba Dental School, Univ Estadual de Campinas, Piracicaba, SP, Brazil.
Braz Oral Res. 2012 Nov-Dec;26(6):550-6. doi: 10.1590/s1806-83242012000600011.
The aim of this study was to evaluate the validity of the bone density value of potential implant sites in HU obtained by a specific cone-beam computed tomography (CBCT) device. In this study, the HU values obtained using a MSCT scanner were used as the gold standard. Twenty mandibles (40 potential implant sites) were scanned using an MSCT scanner (Somatom Sensation 40) and a CBCT scanner (i-CAT). The MSCT images were evaluated using the Syngo CT Workplace software and the CBCT images, using the XoranCat software. The images were evaluated twice by three oral radiologists, at 60 day intervals. The trabecular bone density of the same area was evaluated on both images. Intraclass coefficients (ICC) were calculated to examine the agreement between the examiners and between the two periods of evaluation. The bone density and area of the ROI were compared by the Student t test and Bland-Altman analysis. ICCs were excellent. The mean HU value obtained using CBCT (418.06) was higher than that obtained using MSCT (313.13), with a statistically significant difference (p < 0.0001). In addition, Bland-Altman analysis showed that the HU measures were not equivalent. In conclusion, the bone density in HU with CBCT images obtained using the device studied proved unreliable, since it was higher than that obtained using MSCT.
本研究旨在评估特定锥形束 CT(CBCT)设备获得的 HU 潜在种植体部位骨密度值的有效性。在本研究中,使用 MSCT 扫描仪获得的 HU 值被用作金标准。使用 MSCT 扫描仪(Somatom Sensation 40)和 CBCT 扫描仪(i-CAT)对 20 个下颌骨(40 个潜在种植体部位)进行扫描。使用 Syngo CT Workplace 软件评估 MSCT 图像,使用 XoranCat 软件评估 CBCT 图像。三位口腔放射科医生在 60 天的间隔内对图像进行了两次评估。对同一区域的小梁骨密度进行了评估。计算了组内相关系数(ICC),以检查评估者之间以及两次评估期间的一致性。通过 Student t 检验和 Bland-Altman 分析比较 ROI 的骨密度和面积。ICC 为优。使用 CBCT 获得的平均 HU 值(418.06)高于使用 MSCT 获得的 HU 值(313.13),差异具有统计学意义(p < 0.0001)。此外,Bland-Altman 分析表明 HU 测量值不等效。总之,使用所研究设备获得的 CBCT 图像的 HU 骨密度值不可靠,因为它高于使用 MSCT 获得的 HU 骨密度值。