一种新型 Scheimpflug 和 Placido 盘分析仪测量角膜厚度的精度及其与超声角膜测厚仪的一致性。

Precision of a new Scheimpflug and Placido-disk analyzer in measuring corneal thickness and agreement with ultrasound pachymetry.

机构信息

School of Optometry and Ophthalmology and Eye Hospital, Wenzhou Medical College, Wenzhou, Zhejiang, China.

出版信息

J Cataract Refract Surg. 2013 Feb;39(2):219-24. doi: 10.1016/j.jcrs.2012.10.034. Epub 2012 Dec 3.

Abstract

PURPOSE

To assess the precision of corneal thickness measurements obtained by a new Scheimpflug camera combined with Placido-disk corneal topography (Sirius) and compare the measured values with those obtained by ultrasound (US) pachymetry.

SETTING

Eye Hospital of Wenzhou Medical College, Wenzhou, China.

DESIGN

Comparative evaluation of a diagnostic test or technology.

METHODS

Eyes of healthy subjects were examined with the Scheimpflug-Placido topographer. Central (CCT) and thinnest (TCT) corneal thickness were recorded after 3 consecutive measurements. For US pachymetry, only CCT was measured. Measurements were repeated within 1 week. The within-subject standard deviation (S(w)), test-retest repeatability, coefficient of variation (CoV), and intraclass correlation coefficient (ICC) were calculated to evaluate intrasession repeatability and intersession reproducibility. Interdevice comparison was analyzed with paired t tests and Bland-Altman plots.

RESULTS

The intrasession repeatability of Scheimpflug-Placido measurements was high, with test-retest and CoV close to 9 μm and 0.6% for CCT and TCT, respectively. The intersession test-retest and CoV were close to 10 μm and 0.7%, respectively. The ICC was higher than 0.98 for repeatability and reproducibility. High agreement was found between Scheimpflug-Placido and US pachymetry measurements, with narrow 95% limits of agreement.

CONCLUSIONS

The Scheimpflug-Placido instrument showed excellent intrasession repeatability and intersession reproducibility of CCT and TCT measurements in healthy eyes. High agreement and lack of statistically significant difference suggest that the instrument's TCT and the US pachymetry-CCT measurements can be used interchangeably in subjects with normal corneal thickness.

FINANCIAL DISCLOSURE

No author has a financial or proprietary interest in any material or method mentioned.

摘要

目的

评估新型 Scheimpflug 相机联合 Placido 盘角膜地形图(Sirius)测量角膜厚度的精度,并将测量值与超声(US)角膜测厚术进行比较。

地点

中国温州医学院附属眼视光医院。

设计

诊断测试或技术的比较评估。

方法

对健康受试者的眼睛进行 Scheimpflug-Placido topography 检查。连续测量 3 次后记录中央(CCT)和最薄(TCT)角膜厚度。对于 US 角膜测厚术,仅测量 CCT。在 1 周内重复测量。计算 within-subject standard deviation(S(w))、test-retest repeatability、coefficient of variation(CoV)和 intraclass correlation coefficient(ICC),以评估 intra session 重复性和 inter session 再现性。采用配对 t 检验和 Bland-Altman 图进行 interdevice 比较分析。

结果

Scheimpflug-Placido 测量的 intra session 重复性较高,test-retest 和 CoV 分别接近 9 μm 和 0.6%,用于 CCT 和 TCT。inter session 的 test-retest 和 CoV 分别接近 10 μm 和 0.7%。ICC 重复性和再现性均高于 0.98。Scheimpflug-Placido 和 US 角膜测厚术测量值之间存在高度一致性,95%一致性界限较窄。

结论

Scheimpflug-Placido 仪器在健康眼的 CCT 和 TCT 测量中具有出色的 intra session 重复性和 inter session 再现性。高度一致性和无统计学显著差异表明,该仪器的 TCT 和 US 角膜测厚术-CCT 测量值可在角膜厚度正常的受试者中互换使用。