Suppr超能文献

生物分子分析和负模式探针电喷雾电离在癌症诊断中的应用。

Biomolecular analysis and cancer diagnostics by negative mode probe electrospray ionization.

机构信息

Clean Energy Research Center, University of Yamanashi, 4-3-11 Takeda, Kofu, Yamanashi, 400-8511 Japan.

出版信息

Analyst. 2013 Mar 21;138(6):1682-8. doi: 10.1039/c3an36554a. Epub 2013 Jan 25.

Abstract

We have examined several combinations of solvents and probes with the aim of optimizing the ionization conditions for biomolecules e.g., proteins, peptides and lipids by negative mode probe electrospray ionization mass spectrometry (PESI-MS). With the data presented in this study, negative-mode PESI-MS can be considered as a potential tool for biomolecular analysis and cancer diagnostics because of its simplicity in instrumental configuration. A sharper sampling probe was found to be better for obtaining high quality mass spectra because it can generate stable electrospray without the occurrence of gas breakdown. Although the best conditions may depend on each sample, aqueous organic solvent solutions, especially isopropanol-H(2)O (1/1) with a pH of ≥7, are shown to be preferable for negative-mode PESI-MS, which was successfully applied to colon cancer diagnosis.

摘要

我们已经研究了几种溶剂和探针的组合,旨在通过负离子模式探针电喷雾电离质谱(PESI-MS)优化生物分子(例如蛋白质、肽和脂质)的电离条件。根据本研究提供的数据,由于其仪器配置简单,负离子模式 PESI-MS 可以被认为是生物分子分析和癌症诊断的一种潜在工具。我们发现,更锋利的取样探针更有利于获得高质量的质谱,因为它可以在没有气体击穿的情况下产生稳定的电喷雾。虽然最佳条件可能取决于每个样品,但显示出对于负离子模式 PESI-MS 更有利的是水有机溶剂溶液,特别是异丙醇-H(2)O(1/1),pH 值≥7,其成功应用于结肠癌的诊断。