Suppr超能文献

将活检针作为电极应用于经直肠电阻抗成像。

Incorporating a biopsy needle as an electrode in transrectal electrical impedance imaging.

作者信息

Wan Yuqing, Borsic Andrea, Hartov Alex, Halter Ryan

机构信息

Thayer School of Engineering, Dartmouth College, Hanover, NH 03766, USA.

出版信息

Annu Int Conf IEEE Eng Med Biol Soc. 2012;2012:6220-3. doi: 10.1109/EMBC.2012.6347415.

Abstract

Previous studies have shown that prostate cancer may be detected by a combined transrectal ultrasound and electrical impedance tomography imaging system. However, the sensitivity of the imaging system is limited due to very little current established in the far field distant from the probe surface. Consequently, biopsy needles are introduced to the imaging system to provide current paths in the distal regions. This study demonstrates that image sensitivity can be improved by incorporating the needle electrodes. A phantom experiment is presented to show that contrast to the background is enhanced by 17.4% when imaging with needle electrodes. Simulated reconstructions and some preliminary clinical data also suggest the sensitivity improvement. In summary, TREIT with needle electrodes in the tissue may have great potential in future clinical prostate cancer detection.

摘要

先前的研究表明,经直肠超声和电阻抗断层成像系统相结合可检测出前列腺癌。然而,由于在远离探头表面的远场中建立的电流非常小,成像系统的灵敏度受到限制。因此,将活检针引入成像系统以在远端区域提供电流路径。本研究表明,通过加入针电极可以提高图像灵敏度。进行了一项模型实验,结果表明使用针电极成像时与背景的对比度提高了17.4%。模拟重建和一些初步临床数据也表明了灵敏度的提高。总之,在组织中使用针电极的经直肠电阻抗断层成像在未来临床前列腺癌检测中可能具有巨大潜力。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/3e95/3725641/d0d7ea2fd62b/nihms-496759-f0001.jpg

相似文献

3
Phantom Studies of Fused-Data TREIT Using Only Biopsy-Probe Electrodes.仅使用活检探头电极的融合数据 TREIT 的幻影研究。
IEEE Trans Med Imaging. 2020 Nov;39(11):3367-3378. doi: 10.1109/TMI.2020.2992453. Epub 2020 Oct 28.
8
Electrical impedance endotomography.电阻抗内镜断层成像术。
Phys Med Biol. 2002 Jul 7;47(13):2189-202. doi: 10.1088/0031-9155/47/13/301.
9
Electrical impedance of stainless steel needle electrodes.不锈钢针电极的电阻抗。
Ann Biomed Eng. 2010 Jul;38(7):2371-82. doi: 10.1007/s10439-010-9989-2. Epub 2010 Mar 9.

本文引用的文献

7
Electrical impedance spectroscopy of the human prostate.人体前列腺的电阻抗光谱分析
IEEE Trans Biomed Eng. 2007 Jul;54(7):1321-7. doi: 10.1109/TBME.2007.897331.
9
Electrical impedance endotomography.电阻抗内镜断层成像术。
Phys Med Biol. 2002 Jul 7;47(13):2189-202. doi: 10.1088/0031-9155/47/13/301.