Wan Yuqing, Borsic Andrea, Hartov Alex, Halter Ryan
Thayer School of Engineering, Dartmouth College, Hanover, NH 03766, USA.
Annu Int Conf IEEE Eng Med Biol Soc. 2012;2012:6220-3. doi: 10.1109/EMBC.2012.6347415.
Previous studies have shown that prostate cancer may be detected by a combined transrectal ultrasound and electrical impedance tomography imaging system. However, the sensitivity of the imaging system is limited due to very little current established in the far field distant from the probe surface. Consequently, biopsy needles are introduced to the imaging system to provide current paths in the distal regions. This study demonstrates that image sensitivity can be improved by incorporating the needle electrodes. A phantom experiment is presented to show that contrast to the background is enhanced by 17.4% when imaging with needle electrodes. Simulated reconstructions and some preliminary clinical data also suggest the sensitivity improvement. In summary, TREIT with needle electrodes in the tissue may have great potential in future clinical prostate cancer detection.