Suppr超能文献

使用旋转式 Scheimpflug 相机测量眼内晶状体计算用角膜屈光力的方法比较。

Comparison of methods to measure corneal power for intraocular lens power calculation using a rotating Scheimpflug camera.

机构信息

G.B. Bietti Foundation IRCCS, Rome, Italy.

出版信息

J Cataract Refract Surg. 2013 Apr;39(4):598-604. doi: 10.1016/j.jcrs.2012.11.022. Epub 2013 Feb 9.

Abstract

PURPOSE

To assess the accuracy of corneal power measurements provided by a Scheimpflug camera (Pentacam HR) for intraocular lens (IOL) power calculation in unoperated eyes and compare the results with those of simulated keratometry (SimK) performed with a Placido-disk corneal topographer (Keratron).

SETTING

Private practice.

DESIGN

Evaluation of diagnostic test.

METHODS

Eight Scheimpflug camera corneal power measurements were analyzed: (1) average K, (2) true net power calculated using the Gaussian optics formula, (3) total corneal refractive power at 2.0 mm calculated by ray tracing on a ring and (4) as the average of the zone inside the ring, (5) total corneal refractive power at 3.0 mm on a ring and (6) as the average of the zone inside the ring, (7) the equivalent K reading at 3.0 mm and (8) at 4.5 mm. The IOL power was calculated using the Hoffer Q, Holladay 1, and SRK/T formulas.

RESULTS

No statistically significant differences were observed between any corneal power measurements, including simulated K, in 41 consecutive patients. The latter showed slightly lower mean absolute errors with all 3 formulas (range 0.26 to 0.27 diopter [D]). The Scheimpflug camera gave the lowest median absolute errors with all formulas; that is, the 3.0 mm equivalent K reading with the Hoffer Q formula (0.18 D) and Holladay 1 formula (0.17 D) and the 2.0 mm total corneal refractive power ring with the SRK/T formula (0.18 D).

CONCLUSION

Corneal power measurements provided by the Scheimpflug camera and Placido disk corneal topographer displayed comparable accuracy in IOL power calculation.

摘要

目的

评估在未行手术的眼中,Scheimpflug 相机(Pentacam HR)提供的角膜屈光力测量值用于人工晶状体(IOL)屈光力计算的准确性,并与 Placido 盘角膜地形图仪(Keratron)模拟角膜曲率计(SimK)的结果进行比较。

设置

私人执业。

设计

诊断测试评估。

方法

分析了 8 次 Scheimpflug 相机角膜屈光力测量值:(1)平均 K 值,(2)使用高斯光学公式计算的真实净屈光力,(3)通过在环上进行光线追踪计算的总角膜屈光力 2.0mm,以及(4)作为环内平均值,(5)总角膜屈光力 3.0mm 在环上,以及(6)作为环内平均值,(7)等效 K 值在 3.0mm 处的读数,以及(8)在 4.5mm 处的读数。使用 Hoffer Q、Holladay 1 和 SRK/T 公式计算 IOL 屈光力。

结果

在 41 例连续患者中,任何角膜屈光力测量值,包括模拟角膜曲率计,均未观察到统计学差异。所有 3 种公式的平均绝对误差略低(范围为 0.26 至 0.27 屈光度 [D])。Scheimpflug 相机的所有公式的中位绝对误差均最低;即,Hoffer Q 公式(0.18 D)和 Holladay 1 公式(0.17 D)的 3.0mm 等效 K 值读数以及 SRK/T 公式(0.18 D)的 2.0mm 总角膜屈光力环。

结论

Scheimpflug 相机和 Placido 盘角膜地形图仪提供的角膜屈光力测量值在计算 IOL 屈光力方面具有相当的准确性。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验