Suppr超能文献

乳腺 3D T2 加权自旋回波成像。

3D T2-weighted spin echo imaging in the breast.

机构信息

Department of Radiology, Stanford University, Stanford, California, USA.

出版信息

J Magn Reson Imaging. 2014 Feb;39(2):332-8. doi: 10.1002/jmri.24151. Epub 2013 Apr 17.

Abstract

PURPOSE

To evaluate the performance of 2D versus 3D T2-weighted spin echo imaging in the breast.

MATERIALS AND METHODS

2D and 3D T2-weighted images were acquired in 25 patients as part of a clinically indicated breast magnetic resonance imaging (MRI) exam. Lesion-to-fibroglandular tissue signal ratio was measured in 16 identified lesions. Clarity of lesion morphology was assessed through a blinded review by three radiologists. Instances demonstrating the potential diagnostic contribution of 3D versus 2D T2-weighted imaging in the breast were noted through unblinded review by a fourth radiologist.

RESULTS

The lesion-to-fibroglandular tissue signal ratio was well correlated between 2D and 3D T2-weighted images (R(2)  = 0.93). Clarity of lesion morphology was significantly better with 3D T2-weighted imaging for all observers based on a McNemar test (P ≤ 0.02, P ≤ 0.01, P ≤ 0.03). Instances indicating the potential diagnostic contribution of 3D T2-weighted imaging included improved depiction of signal intensity and improved alignment between DCE and T2-weighted findings.

CONCLUSION

In this pilot study, 3D T2-weighted imaging provided comparable contrast and improved depiction of lesion morphology in the breast in comparison to 2D T2-weighted imaging. Based on these results further investigation to determine the diagnostic impact of 3D T2-weighted imaging in breast MRI is warranted.

摘要

目的

评估二维与三维 T2 加权自旋回波成像在乳腺中的性能。

材料与方法

在 25 例临床疑似乳腺磁共振成像(MRI)检查的患者中采集二维和三维 T2 加权图像。在 16 个已识别的病变中测量病变与纤维腺体组织的信号比。三位放射科医生通过盲法评估评估病变形态的清晰度。通过第四位放射科医生的非盲法评估,记录了在乳腺中 3D 与 2D T2 加权成像潜在诊断价值的实例。

结果

二维和三维 T2 加权图像之间的病变与纤维腺体组织的信号比具有很好的相关性(R²=0.93)。基于 McNemar 检验,所有观察者均认为三维 T2 加权成像对病变形态的清晰度明显更好(P≤0.02、P≤0.01、P≤0.03)。表明 3D T2 加权成像具有潜在诊断价值的实例包括改善了信号强度的显示和 DCE 与 T2 加权成像结果之间的一致性。

结论

在这项初步研究中,与二维 T2 加权成像相比,三维 T2 加权成像在乳腺中提供了可比的对比度,并改善了病变形态的显示。基于这些结果,需要进一步研究确定 3D T2 加权成像在乳腺 MRI 中的诊断价值。