Moylan Rebecca, Aland Trent, Kairn Tanya
Premion, Chermside Medical Complex, 956 Gympie Road, Chermside, QLD 4032, Australia.
Australas Phys Eng Sci Med. 2013 Sep;36(3):331-7. doi: 10.1007/s13246-013-0206-0. Epub 2013 Jun 26.
Radiochromic film has the potential to provide accurate in vivo dosimetry measurements. However, it is not known whether small film pieces can still provide accurate dosimetric results. The use of small film pieces is of particular interest in regions of interest (ROIs) such as the eye, or where the patient's contour changes rapidly. This study examines the dosimetric accuracy of Gafchromic EBT2 and EBT3 models of radiochromic film and its dependence on film size, ROI size, and height above the scan bed for 6 MV photons and 9 MeV electrons. Films cut to sizes of 5.0 × 5.0, 10.0 × 10.0, 20.0 × 20.0, and 40.0 × 40.0 mm² were tested and it was found that there was no increase in uncertainty of dose when even the smallest film sizes were used. For a film 5.0 × 5.0 mm², ROIs of 1.4 × 1.4, 2.1 × 2.1 and 3.5 × 3.5 mm² were tested and it was found that the ROI size of 2.1 × 2.1 mm² was the most accurate. The standard deviation of the EBT3 placed on the glass (2.1%) was larger than the standard deviation of the EBT3 film raised above the glass (1.2%), therefore it is recommended that film is scanned raised above the scan bed. The general dosimetric performance of EBT3 was comparable to EBT2. We conclude that film pieces as small as 5.0 × 5.0 mm² could be used for the purpose of in vivo dosimetry of radiotherapy treatments.
放射变色薄膜有潜力提供准确的体内剂量测定。然而,尚不清楚小薄膜片是否仍能提供准确的剂量测定结果。在诸如眼睛等感兴趣区域(ROI),或患者轮廓快速变化的部位,使用小薄膜片尤其有意义。本研究考察了Gafchromic EBT2和EBT3型号放射变色薄膜的剂量测定准确性,以及其对薄膜尺寸、感兴趣区域尺寸和高于扫描床高度的依赖性,针对6兆伏光子和9兆电子伏电子进行了研究。测试了切割成5.0×5.0、10.0×10.0、20.0×20.0和40.0×40.0平方毫米尺寸的薄膜,发现即使使用最小尺寸的薄膜,剂量不确定性也没有增加。对于一片5.0×5.0平方毫米的薄膜,测试了尺寸为1.4×1.4、2.1×2.1和3.5×3.5平方毫米的感兴趣区域,发现尺寸为2.1×2.1平方毫米的感兴趣区域最准确。放置在玻璃上的EBT3的标准偏差(2.1%)大于高于玻璃的EBT3薄膜的标准偏差(1.2%),因此建议在高于扫描床的位置扫描薄膜。EBT3的总体剂量测定性能与EBT2相当。我们得出结论,小至5.0×5.0平方毫米的薄膜片可用于放射治疗的体内剂量测定。