St. Thomas' Hospital, Lambeth Palace Road, London, SE1 7EH, United Kingdom.
Indian J Ophthalmol. 2013 Nov;61(11):653-8. doi: 10.4103/0301-4738.117804.
Lack of clinical tests to quantify spatial components of distortion in patients with full thickness macular holes (FTMH) and epiretinal membranes (ERM).
To develop a test for subjective evaluation of visual distortion in the central visual field around fixation in patients with unilateral FTMH or ERM.
Prospective case-control study carried out at tertiary referral center.
Twenty-five patients with unilateral macular disease (13 macular epiretinal membranes, 12 full-thickness macular holes), and nine controls (without ocular pathology) underwent ophthalmological examination with logMAR ETDRS visual acuity, near vision and contrast sensitivity assessed. Macular optical coherence tomography and metamorphopsia assessment using Morphision test was also carried out. This test consists of a set of modified Amsler charts for detection, identification, and subjective quantification of visual distortion in the central visual field around fixation. Morphision test content and construct validity, and reliability (test-retest method) were evaluated. Sixteen patients completed an unstructured survey on test performance and preference.
Every patient with unilateral FTMH or ERM identified a particular chart using Morphision test (content validity). None of the normal subjects without symptoms of metamorphopsia identified any distortion (construct validity). Test-retest showed a 100% consistency for frequency and 67% for amplitude. The mean amplitude difference between measurements was 0.02 degrees (SD = 0.038). The coefficient of repeatability was 0.075. There was a correlation between Morphision amplitude score and visual acuity and contrast sensitivity, individually.
Morphision test allowed detection and subjective quantification of metamorphopsia in the clinical setting in our patients with unilateral macular epiretinal membranes and full thickness macular holes.
缺乏用于量化全层黄斑裂孔 (FTMH) 和视网膜前膜 (ERM) 患者失真空间成分的临床测试。
开发一种用于评估单侧 FTMH 或 ERM 患者固定注视中心视野中视觉扭曲的主观测试方法。
在三级转诊中心进行的前瞻性病例对照研究。
25 名单侧黄斑疾病患者(13 名黄斑视网膜前膜,12 名全层黄斑裂孔)和 9 名对照者(无眼部病变)接受眼科检查,包括 logMAR ETDRS 视力、近视力和对比敏感度评估。还进行了黄斑光学相干断层扫描和 Morphision 测试评估的形觉错觉评估。该测试由一组用于检测、识别和主观量化固定注视中心视觉场中视觉扭曲的改良 Amsler 图表组成。评估了 Morphision 测试的内容和结构有效性以及可靠性(测试-重测法)。16 名患者完成了一项关于测试性能和偏好的非结构化调查。
使用 Morphision 测试,每位单侧 FTMH 或 ERM 患者均能识别出特定图表(内容有效性)。没有任何无症状的变形症患者能识别出任何变形(结构有效性)。测试-重测显示频率的一致性为 100%,幅度的一致性为 67%。两次测量之间的平均幅度差异为 0.02 度(SD=0.038)。可重复性系数为 0.075。Morphision 幅度评分与视力和对比敏感度之间存在相关性。
Morphision 测试允许在单侧黄斑视网膜前膜和全层黄斑裂孔患者的临床环境中检测和主观量化变形症。