Suppr超能文献

飞秒激光(IntraLase FS150、WaveLight FS200)与机械角膜刀制作 LASIK 瓣基质床的对比研究。

Comparative study of stromal bed of LASIK flaps created with femtosecond lasers (IntraLase FS150, WaveLight FS200) and mechanical microkeratome.

机构信息

Vardinoyiannion Eye Institute of Crete (VEIC), Faculty of Medicine, University of Crete, , Heraklion, Crete, Greece.

出版信息

Br J Ophthalmol. 2014 Jan;98(1):133-7. doi: 10.1136/bjophthalmol-2013-304023. Epub 2013 Nov 1.

Abstract

BACKGROUND/AIM: To compare stromal bed smoothness in LASIK flaps created with two different femtosecond lasers (IntraLase FS150 and WaveLight FS200) and a mechanical microkeratome (MMK) (Carriazo-Pendular microkeratome).

METHODS

Sixty freshly enucleated porcine eyes were used for the study. Twenty flaps were created and constitute each of the following three groups: IntraLase FS150 (IFS group), WaveLight FS200 (WFS group) and MMK (MMK group). In each of the three groups, 10 flaps were created with intended thickness of 110 μm and another 10 flaps with 130 μm. Images were assessed with light microscopy and scanning electron microscopy. Qualitative surface roughness grading of the images was performed by five masked observers and statistical comparisons of scores were made between groups.

RESULTS

The results of qualitative surface roughness grading demonstrated that there was no significant difference between the two femtosecond (FS) lasers when comparing the subgroups with the same flap thickness (p>0.05 in both comparisons, Mann-Whitney U test). In addition, there was no statistically significant difference (p>0.05) in each flap maker group between different thickness subgroups. However, the scores of the FS laser subgroups were significantly better than the scores of the MMK subgroups with corresponding flap thicknesses (p<0.05, Mann-Whitney U test).

CONCLUSIONS

Our study demonstrates that the IntraLase FS150 and the WaveLight FS200 had a similar performance and provided flaps with smooth surfaces. In comparison to the MMK, the studied femtosecond laser systems had a superior performance in terms of smoothness.

摘要

背景/目的:比较两种不同的飞秒激光(IntraLase FS150 和 WaveLight FS200)和机械角膜刀(MMK)(Carriazo-Pendular 微角膜刀)制作的 LASIK 瓣的基质床平滑度。

方法

本研究使用了 60 只新鲜取出的猪眼。每组 20 个瓣,分为以下三组:IntraLase FS150(IFS 组)、WaveLight FS200(WFS 组)和 MMK(MMK 组)。在每组中,10 个瓣的预设厚度为 110μm,另外 10 个瓣的预设厚度为 130μm。用光学显微镜和扫描电子显微镜评估图像。5 位蒙面观察者对图像进行定性表面粗糙度分级,并对组间评分进行统计学比较。

结果

定性表面粗糙度分级结果表明,当比较相同瓣厚度的亚组时,两种飞秒(FS)激光之间没有显著差异(两种比较的 p 值均>0.05,Mann-Whitney U 检验)。此外,在每个瓣制作器组中,不同厚度亚组之间没有统计学差异(p 值均>0.05)。然而,FS 激光亚组的评分明显优于相应瓣厚度的 MMK 亚组(p 值均<0.05,Mann-Whitney U 检验)。

结论

我们的研究表明,IntraLase FS150 和 WaveLight FS200 具有相似的性能,提供了表面光滑的瓣。与 MMK 相比,研究中的飞秒激光系统在平滑度方面表现更优。