Doern Christopher D, Mirrett Stanley, Halstead Diane, Abid Joan, Okada Pamela, Reller L Barth
Department of Pathology, Virginia Commonwealth University Health System, Richmond, Virginia, USA
Clinical Microbiology Laboratory, Duke University Medical Center, Durham, North Carolina, USA.
J Clin Microbiol. 2014 Jun;52(6):1898-900. doi: 10.1128/JCM.00175-14. Epub 2014 Mar 19.
We conducted a controlled clinical comparison of PF Plus, the new pediatric medium with adsorbent polymeric beads, versus the charcoal-containing PF medium in the BacT/Alert blood culture system. A total of 2,381 pediatric cultures were enrolled, and 1,703 (71.5%) were deemed to be compliant and acceptable for analysis. Seventy-two cultures (4.2%) had a positive result with 80 clinically significant microorganisms. The results showed that the PF Plus medium yielded more clinically significant microorganisms than the BacT/Alert system (P < 0.05). In addition, the PF Plus bottle yielded positive results an average of 5.0 h earlier than the PF bottle for compliant clinically significant cultures (18.3 h versus 23.2 h, P = 0.004). PF Plus is an improved medium for detecting microorganisms that cause pediatric bloodstream infections.
我们在BacT/Alert血培养系统中,对含吸附性聚合物微珠的新型儿科培养基PF Plus与含活性炭的PF培养基进行了对照临床比较。共纳入2381份儿科培养样本,其中1703份(71.5%)被认为符合要求且可用于分析。72份培养样本(4.2%)检测呈阳性,分离出80种具有临床意义的微生物。结果显示,PF Plus培养基分离出的具有临床意义的微生物比BacT/Alert系统更多(P < 0.05)。此外,对于符合要求且具有临床意义的培养样本,PF Plus瓶平均比PF瓶早5.0小时得出阳性结果(分别为18.3小时和23.2小时,P = 0.004)。PF Plus是一种用于检测引起儿科血流感染微生物的改良培养基。