离子淌度质谱提高同位素质谱标签定量分析。

Improved isobaric tandem mass tag quantification by ion mobility mass spectrometry.

机构信息

Department of Chemistry, University of Wisconsin-Madison.

School of Pharmacy, University of Wisconsin-Madison.

出版信息

Rapid Commun Mass Spectrom. 2014 May 15;28(9):1051-1060. doi: 10.1002/rcm.6875.

Abstract

RATIONALE

Isobaric tandem mass tags are an attractive alternative to mass difference tags and label-free approaches for quantitative proteomics due to the high degree of multiplexing that can be performed with their implementation. A drawback of tandem mass tags are that the co-isolation and co-fragmentation of labeled peptide precursors can result in chimeric tandem mass (MS/MS) spectra that can underestimate the fold-change expression of each peptide. Ion mobility (IM) separations coupled to quadrupole time-of-flight (Q-TOF) instruments have the potential to mitigate MS/MS spectra chimeracy since IM-MS has the ability to separate ions based on charge, m/z, and collision cross section (CCS).

METHODS

Two complex protein mixtures, labeled with DiLeu isobaric tandem mass tags in opposite ratios, were mixed together and analyzed by data-dependent LC/IM-MS/MS. The accuracy of reporters from interfering pairs was compared with and without IM separation.

RESULTS

IM separation was able to mitigate isobaric interference from differentially charged interfering ion pairs, as well as pairs of the same charge. Of the eight example precursors shown, only one had reporters that remained compressed below the significance threshold after IM separation.

CONCLUSIONS

The results of this investigation demonstrate proof-of-principle that IM separation of tagged precursors prior to MS/MS fragmentation can help mitigate quantitative inaccuracies caused by isobaric interference. Future improvements of the method would include software for automated correction and use of higher resolution IM instrumentations.

摘要

原理

由于串联质谱标签的实施可以实现高度的多重化,因此与质量差异标签和无标记方法相比,串联质谱标签是定量蛋白质组学的一种有吸引力的选择。串联质谱标签的一个缺点是标记的肽前体的共分离和共碎裂可能导致嵌合串联质谱 (MS/MS) 谱,从而低估每个肽的倍数变化表达。离子淌度 (IM) 与四极杆飞行时间 (Q-TOF) 仪器的分离具有减轻 MS/MS 谱嵌合的潜力,因为 IM-MS 具有根据电荷、m/z 和碰撞截面 (CCS) 分离离子的能力。

方法

两种复杂的蛋白质混合物用 DiLeu 等摩尔串联质谱标签以相反的比例标记,然后混合在一起,并通过依赖数据的 LC/IM-MS/MS 进行分析。比较了有无 IM 分离时干扰对的报告器的准确性。

结果

IM 分离能够减轻带不同电荷的干扰离子对以及相同电荷的离子对的等摩尔干扰。在所显示的 8 个示例前体中,只有一个具有在 IM 分离后报告器仍低于显着性阈值的报告器。

结论

该研究的结果证明了在 MS/MS 碎裂之前对标记前体进行 IM 分离有助于减轻等摩尔干扰引起的定量不准确的原理证明。该方法的未来改进将包括用于自动校正的软件和使用更高分辨率的 IM 仪器。

相似文献

引用本文的文献

3
Quantitative Peptidomics: General Considerations.定量肽组学:一般考虑因素。
Methods Mol Biol. 2024;2758:89-108. doi: 10.1007/978-1-0716-3646-6_5.
7
Chemical isotope labeling for quantitative proteomics.化学同位素标记定量蛋白质组学。
Mass Spectrom Rev. 2023 Mar;42(2):546-576. doi: 10.1002/mas.21709. Epub 2021 Jun 6.
9
A New Quantification Method Using Electrochemical Mass Spectrometry.电化学质谱法的一种新定量方法。
J Am Soc Mass Spectrom. 2019 Apr;30(4):685-693. doi: 10.1007/s13361-018-2116-6. Epub 2019 Jan 2.
10
Sample Multiplexing Strategies in Quantitative Proteomics.定量蛋白质组学中的样本多路复用策略
Anal Chem. 2019 Jan 2;91(1):178-189. doi: 10.1021/acs.analchem.8b05626. Epub 2018 Dec 18.

本文引用的文献

9
Enhancements in travelling wave ion mobility resolution. travelling wave 离子迁移率分辨率的提高。
Rapid Commun Mass Spectrom. 2011 Jun 15;25(11):1559-66. doi: 10.1002/rcm.5013.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验