Inness Emma K, Moutrie Vaughan, Charles Paul H
Radiation Oncology, Princess Alexandra Hospital, Woolloongabba, Brisbane, QLD, 4102, Australia,
Australas Phys Eng Sci Med. 2014 Jun;37(2):385-91. doi: 10.1007/s13246-014-0272-y. Epub 2014 Apr 24.
A computed tomography number to relative electron density (CT-RED) calibration is performed when commissioning a radiotherapy CT scanner by imaging a calibration phantom with inserts of specified RED and recording the CT number displayed. In this work, CT-RED calibrations were generated using several commercially available phantoms to observe the effect of phantom geometry on conversion to electron density and, ultimately, the dose calculation in a treatment planning system. Using an anthropomorphic phantom as a gold standard, the CT number of a material was found to depend strongly on the amount and type of scattering material surrounding the volume of interest, with the largest variation observed for the highest density material tested, cortical bone. Cortical bone gave a maximum CT number difference of 1,110 when a cylindrical insert of diameter 28 mm scanned free in air was compared to that in the form of a 30 × 30 cm(2) slab. The effect of using each CT-RED calibration on planned dose to a patient was quantified using a commercially available treatment planning system. When all calibrations were compared to the anthropomorphic calibration, the largest percentage dose difference was 4.2 % which occurred when the CT-RED calibration curve was acquired with heterogeneity inserts removed from the phantom and scanned free in air. The maximum dose difference observed between two dedicated CT-RED phantoms was ±2.1 %. A phantom that is to be used for CT-RED calibrations must have sufficient water equivalent scattering material surrounding the heterogeneous objects that are to be used for calibration.
在调试放射治疗CT扫描仪时,通过对带有特定相对电子密度(RED)插入物的校准体模进行成像并记录显示的CT值,来进行计算机断层扫描数值到相对电子密度(CT-RED)的校准。在这项工作中,使用几种市售体模生成CT-RED校准,以观察体模几何形状对转换为电子密度的影响,并最终观察其对治疗计划系统中剂量计算的影响。以拟人化体模作为金标准,发现一种材料的CT值强烈依赖于感兴趣体积周围散射材料的数量和类型,在所测试的最高密度材料(皮质骨)中观察到的变化最大。当将直径28 mm的圆柱形插入物在空气中自由扫描与30×30 cm²平板形式的插入物进行比较时,皮质骨的最大CT值差异为1110。使用市售治疗计划系统对使用每种CT-RED校准对患者计划剂量的影响进行了量化。当将所有校准与拟人化校准进行比较时,最大百分比剂量差异为4.2%,这发生在从体模中移除异质性插入物并在空气中自由扫描获取CT-RED校准曲线时。在两个专用CT-RED体模之间观察到的最大剂量差异为±2.1%。用于CT-RED校准的体模必须在用于校准的异质物体周围有足够的水等效散射材料。