¹⁹²铱近距离治疗源在不同介质中的剂量分布。

Dose distributions of an ¹⁹²Ir brachytherapy source in different media.

作者信息

Wu C H, Liao Y J, Liu Y W Hsueh, Hung S K, Lee M S, Hsu S M

机构信息

Institute of Nuclear Engineering and Science, National Tsing Hua University, Hsinchu 300, Taiwan.

School of Medical Laboratory Science and Biotechnology, College of Medical Science and Technology, Taipei Medical University, Taipei 110, Taiwan.

出版信息

Biomed Res Int. 2014;2014:946213. doi: 10.1155/2014/946213. Epub 2014 Apr 7.

Abstract

This study used MCNPX code to investigate the brachytherapy (192)Ir dose distributions in water, bone, and lung tissue and performed radiophotoluminescent glass dosimeter measurements to verify the obtained MCNPX results. The results showed that the dose-rate constant, radial dose function, and anisotropy function in water were highly consistent with data in the literature. However, the lung dose near the source would be overestimated by up to 12%, if the lung tissue is assumed to be water, and, hence, if a tumor is located in the lung, the tumor dose will be overestimated, if the material density is not taken into consideration. In contrast, the lung dose far from the source would be underestimated by up to 30%. Radial dose functions were found to depend not only on the phantom size but also on the material density. The phantom size affects the radial dose function in bone more than those in the other tissues. On the other hand, the anisotropy function in lung tissue was not dependent on the radial distance. Our simulation results could represent valid clinical reference data and be used to improve the accuracy of the doses delivered during brachytherapy applied to patients with lung cancer.

摘要

本研究使用MCNPX代码研究近距离放射治疗中(192)Ir在水、骨和肺组织中的剂量分布,并进行了放射性光致发光玻璃剂量计测量以验证所获得的MCNPX结果。结果表明,水中的剂量率常数、径向剂量函数和各向异性函数与文献中的数据高度一致。然而,如果将肺组织假定为水,那么源附近的肺剂量将被高估高达12%,因此,如果肿瘤位于肺中,若不考虑材料密度,肿瘤剂量将被高估。相反,远离源的肺剂量将被低估高达30%。发现径向剂量函数不仅取决于模体尺寸,还取决于材料密度。模体尺寸对骨中径向剂量函数的影响大于对其他组织的影响。另一方面,肺组织中的各向异性函数不依赖于径向距离。我们的模拟结果可代表有效的临床参考数据,并用于提高应用于肺癌患者的近距离放射治疗期间所输送剂量的准确性。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/7989/3997138/21bce84c0324/BMRI2014-946213.001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验