曝光因素对 CBCT 体素值变异性的影响:一项体模研究。

Influence of exposure factors on the variability of CBCT voxel values: a phantom study.

机构信息

Department of Oral Diagnosis, Division of Oral Radiology, State University of Campinas, Piracicaba Dental School, Piracicaba, Sao Paulo, Brazil.

出版信息

Dentomaxillofac Radiol. 2014;43(6):20140128. doi: 10.1259/dmfr.20140128. Epub 2014 May 28.

Abstract

OBJECTIVES

To assess the influence of milliamperage and kilovolt peak (kVp) on the variability of cone beam CT (CBCT) voxel values.

METHODS

CBCT scans were obtained from radiographic phantoms in varying concentrations of dipotassium hydrogen phosphate solutions (200-1200 mg ml(-1)) under different protocols of milliamperage and kVp. In addition, scans were performed with and without a dental implant and exo-mass. The variability of CBCT voxel values was measured on each scan, and factorial analysis of variance and the post hoc Tukey test were performed (α = 0.05). Linear regression was performed to assess the relationship between voxel value variability and dipotassium hydrogen phosphate concentration.

RESULTS

milliamperage and the presence of a dental implant did not produce significant interference (p = 0.28 and 0.87, respectively) in voxel value variability. Scans at the highest kVp value presented a significant reduction (p ≤ 0.0001) in voxel value variability when only exo-mass was not present. Voxel value variability was not influenced by exo-mass in scans at the highest levels of milliamperage and kVp. The presence of exo-mass produced a significant reduction (p ≤ 0.0001) in voxel value variability in most of the scans. Higher concentrations yielded greater variations in voxel values in all scans, except for those operating at the highest levels of mAs and kVp.

CONCLUSIONS

mAs did not influence the variability of CBCT voxel values; higher kVp reduced such variability when only the object was smaller than the field of view.

摘要

目的

评估毫安和千伏峰值(kVp)对锥形束 CT(CBCT)体素值变异性的影响。

方法

在不同毫安和 kVp 协议下,从不同浓度的磷酸二氢钾溶液(200-1200mg/ml)的放射影像体模中获得 CBCT 扫描。此外,还在有无牙种植体和外源性质量的情况下进行了扫描。在每次扫描上测量 CBCT 体素值的变异性,并进行析因方差分析和事后 Tukey 检验(α=0.05)。进行线性回归以评估体素值变异性与磷酸二氢钾浓度之间的关系。

结果

毫安和牙种植体的存在并未对体素值变异性产生显著干扰(p=0.28 和 0.87)。当仅不存在外源性质量时,最高 kVp 值的扫描会导致体素值变异性显著降低(p≤0.0001)。在最高毫安和 kVp 值的扫描中,外源性质量不会影响体素值变异性。外源性质量的存在会导致大多数扫描中的体素值变异性显著降低(p≤0.0001)。除了在最高毫安和 kVp 水平下的扫描外,较高的浓度会导致所有扫描中的体素值变化更大。

结论

毫安不会影响 CBCT 体素值的变异性;当物体小于视野时,较高的 kVp 会降低这种变异性。

相似文献

3
Influence of anatomical location on CT numbers in cone beam computed tomography.解剖位置对锥形束 CT 中 CT 数的影响。
Oral Surg Oral Med Oral Pathol Oral Radiol. 2013 Apr;115(4):558-64. doi: 10.1016/j.oooo.2013.01.021.
4
Influence of voxel size on cone beam computed tomography artifacts arising from the exomass.体素大小对源于外生骨块的锥形束计算机断层扫描伪影的影响。
Oral Surg Oral Med Oral Pathol Oral Radiol. 2021 Oct;132(4):456-464. doi: 10.1016/j.oooo.2020.12.003. Epub 2021 Jan 6.
5
Metallic materials in the exomass impair cone beam CT voxel values.外生质量中的金属材料会降低锥形束 CT 体素值。
Dentomaxillofac Radiol. 2018 Jul;47(6):20180011. doi: 10.1259/dmfr.20180011. Epub 2018 Mar 27.

引用本文的文献

本文引用的文献

7
Influence of anatomical location on CT numbers in cone beam computed tomography.解剖位置对锥形束 CT 中 CT 数的影响。
Oral Surg Oral Med Oral Pathol Oral Radiol. 2013 Apr;115(4):558-64. doi: 10.1016/j.oooo.2013.01.021.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验