Suppr超能文献

一次性泪液葡萄糖生物传感器 - 第4部分:评估疗效、安全性和可行性的初步动物模型研究

A Disposable Tear Glucose Biosensor-Part 4: Preliminary Animal Model Study Assessing Efficacy, Safety, and Feasibility.

作者信息

La Belle Jeffrey T, Engelschall Erica, Lan Kenneth, Shah Pankti, Saez Neil, Maxwell Stephanie, Adamson Teagan, Abou-Eid Michelle, McAferty Kenyon, Patel Dharmendra R, Cook Curtiss B

机构信息

Biodesign Institute, Arizona State University, Tempe, AZ, USA Harrington Program of Biomedical Engineering, School of Biological and Health Systems Engineering, Arizona State University, Tempe, AZ, USA

Biodesign Institute, Arizona State University, Tempe, AZ, USA Harrington Program of Biomedical Engineering, School of Biological and Health Systems Engineering, Arizona State University, Tempe, AZ, USA.

出版信息

J Diabetes Sci Technol. 2014 Jan;8(1):109-116. doi: 10.1177/1932296813511741. Epub 2014 Jan 1.

Abstract

OBJECTIVE

A prototype tear glucose (TG) sensor was tested in New Zealand white rabbits to assess eye irritation, blood glucose (BG) and TG lag time, and correlation with BG.

METHODS

A total of 4 animals were used. Eye irritation was monitored by Lissamine green dye and analyzed using image analysis software. Lag time was correlated with an oral glucose load while recording TG and BG readings. Correlation between TG and BG were plotted against one another to form a correlation diagram, using a Yellow Springs Instrument (YSI) and self-monitoring of blood glucose as the reference measurements. Finally, TG levels were calculated using analytically derived expressions.

RESULTS

From repeated testing carried over the course of 12 months, little to no eye irritation was detected. TG fluctuations over time visually appeared to trace the same pattern as BG with an average lag times of 13 minutes. TG levels calculated from the device current measurements ranged from 4 to 20 mg/dL and correlated linearly with BG levels of 75-160 mg/dL (TG = 0.1723 BG = 7.9448 mg/dL; R = .7544).

CONCLUSION

The first steps were taken toward preliminary development of a sensor for self-monitoring of tear glucose (SMTG). No conjunctival irritation in any of the animals was noted. Lag time between TG and BG was found to be noticeable, but a quantitative modeling to correlate lag time in this study is unnecessary. Measured currents from the sensors and the calculated TG showed promising correlation to BG levels. Previous analytical bench marking showed BG and TG levels consistent with other literature.

摘要

目的

在新西兰白兔身上测试一种泪液葡萄糖(TG)传感器原型,以评估眼部刺激性、血糖(BG)和TG滞后时间,以及与BG的相关性。

方法

总共使用了4只动物。通过丽丝胺绿染料监测眼部刺激情况,并使用图像分析软件进行分析。在记录TG和BG读数的同时,将滞后时间与口服葡萄糖负荷相关联。使用黄泉仪器(YSI)和自我血糖监测作为参考测量,将TG和BG之间的相关性相互绘制以形成相关图。最后,使用分析推导的表达式计算TG水平。

结果

在12个月的过程中进行的重复测试显示,几乎未检测到眼部刺激。随着时间的推移,TG波动在视觉上似乎与BG遵循相同的模式,平均滞后时间为13分钟。根据设备电流测量计算出的TG水平范围为4至20mg/dL,并且与75 - 160mg/dL的BG水平呈线性相关(TG = 0.1723 BG - 7.9448mg/dL;R = 0.7544)。

结论

朝着泪液葡萄糖自我监测(SMTG)传感器的初步开发迈出了第一步。未观察到任何动物有结膜刺激。发现TG和BG之间的滞后时间很明显,但在本研究中对滞后时间进行定量建模没有必要。传感器测量的电流和计算出的TG与BG水平显示出有前景的相关性。先前的分析基准表明BG和TG水平与其他文献一致。

相似文献

本文引用的文献

5
Medical device recalls and the FDA approval process.医疗器械召回与美国食品药品监督管理局(FDA)的审批流程。