Department of Ophthalmology, Selcuklu Faculty of Medicine, Selcuk University, Konya, Turkey.
Cornea. 2012 Nov;31(11):1278-81. doi: 10.1097/ICO.0b013e31823f7701.
To compare the accuracy of optical low-coherence reflectometry (Lenstar LS 900; Haag Streit, Köniz, Switzerland) and conventional ultrasound pachymetry (UP) (US 4000; Nidek, Japan) for measuring central corneal thickness and to assess the interexaminer reproducibilities of both devices.
Department of Ophthalmology, Selcuklu Faculty of Medicine, Selcuk University was the study setting. Central corneal thickness (CCT) was measured in both eyes of 65 emmetropic patients by the same examiner using Lenstar LS 900 noncontact biometry and UP. To assess the interexaminer reproducibility, 2 different examiners consecutively obtained CCT measurements in 16 eyes with Lenstar LS 900 noncontact biometry and UP.
Mean CCT measurements with Lenstar LS 900 noncontact biometry and ultrasound biometry were 541.09 ± 33.8 and 545.69 ± 34.0, respectively. Significant correlation was present between these 2 measurements; intraclass correlation coefficient was found to be 95.8% (95% confidence interval: 93.2%-97.4%)(P < 0.001). Interexaminer reproducibility with Lenstar was high, and the intraclass correlation was found to be 99.3% (95% confidence interval: 98.0%-99.7%). We have also investigated the interexaminer reproducibility with UP in these 16 patients, and the results showed a 99.2% (95% confidence interval: 97.7%-99.7%) intraclass correlation.
The results of Lenstar LS 900 noncontact biometry correlated with the results of UP. It may be used as an alternative method to UP for measuring CCT. Because of the mild underestimation of CCT by Lenstar, the measurements should be carefully evaluated in patients whose intraocular pressure must strictly be monitored.
比较光学低相干反射仪(LS 900;Haag Streit,Köniz,瑞士)和传统超声角膜测厚仪(UP)(US 4000;Nidek,日本)测量中央角膜厚度的准确性,并评估这两种设备的检查者间可重复性。
该研究于塞尔丘克大学塞尔柱鲁克医学院眼科进行。同一位检查者使用 LS 900 非接触式生物测量仪和 UP 分别测量 65 例正视眼患者双眼的中央角膜厚度(CCT)。为了评估检查者间的可重复性,2 位不同的检查者使用 LS 900 非接触式生物测量仪和 UP 连续测量 16 只眼的 CCT。
LS 900 非接触式生物测量仪和超声生物测量仪测量的平均 CCT 分别为 541.09 ± 33.8 和 545.69 ± 34.0。这两种测量方法之间存在显著相关性;组内相关系数为 95.8%(95%置信区间:93.2%-97.4%)(P < 0.001)。LS 900 的检查者间可重复性较高,组内相关系数为 99.3%(95%置信区间:98.0%-99.7%)。我们还在这 16 例患者中研究了 UP 的检查者间可重复性,结果显示组内相关系数为 99.2%(95%置信区间:97.7%-99.7%)。
LS 900 非接触式生物测量仪的结果与 UP 的结果相关。它可作为 UP 测量 CCT 的替代方法。由于 LS 900 会轻微低估 CCT,因此在需要严格监测眼内压的患者中,应仔细评估测量结果。