Kim Sang Joon, Choi Choong Gon, Kim Jeong Kon, Yun Sung-Cheol, Jahng Geon-Ho, Jeong Ha-Kyu, Kim Eun Ju
Department of Radiology, Asan Medical Center, University of Ulsan College of Medicine, Seoul 138-736, Korea.
Department of Biostatistics, University of Ulsan College of Medicine, Seoul 138-736, Korea.
Korean J Radiol. 2015 Mar-Apr;16(2):297-303. doi: 10.3348/kjr.2015.16.2.297. Epub 2015 Feb 27.
To validate the usefulness of a diffusional anisotropic capillary array phantom and to investigate the effects of diffusion tensor imaging (DTI) parameter changes on diffusion fractional anisotropy (FA) and apparent diffusion coefficient (ADC) using the phantom.
Diffusion tensor imaging of a capillary array phantom was performed with imaging parameter changes, including voxel size, number of sensitivity encoding (SENSE) factor, echo time (TE), number of signal acquisitions, b-value, and number of diffusion gradient directions (NDGD), one-at-a-time in a stepwise-incremental fashion. We repeated the entire series of DTI scans thrice. The coefficients of variation (CoV) were evaluated for FA and ADC, and the correlation between each MR imaging parameter and the corresponding FA and ADC was evaluated using Spearman's correlation analysis.
The capillary array phantom CoVs of FA and ADC were 7.1% and 2.4%, respectively. There were significant correlations between FA and SENSE factor, TE, b-value, and NDGD, as well as significant correlations between ADC and SENSE factor, TE, and b-value.
A capillary array phantom enables repeated measurements of FA and ADC. Both FA and ADC can vary when certain parameters are changed during diffusion experiments. We suggest that the capillary array phantom can be used for quality control in longitudinal or multicenter clinical studies.
验证扩散各向异性毛细管阵列体模的实用性,并使用该体模研究扩散张量成像(DTI)参数变化对扩散分数各向异性(FA)和表观扩散系数(ADC)的影响。
对毛细管阵列体模进行扩散张量成像,每次逐一逐步增加成像参数,包括体素大小、敏感度编码(SENSE)因子数量、回波时间(TE)、信号采集次数、b值和扩散梯度方向数量(NDGD)。我们将整个DTI扫描系列重复三次。评估FA和ADC的变异系数(CoV),并使用Spearman相关分析评估每个磁共振成像参数与相应的FA和ADC之间的相关性。
毛细管阵列体模FA和ADC的CoV分别为7.1%和2.4%。FA与SENSE因子、TE、b值和NDGD之间存在显著相关性,ADC与SENSE因子、TE和b值之间也存在显著相关性。
毛细管阵列体模能够对FA和ADC进行重复测量。在扩散实验中,当某些参数改变时,FA和ADC都会发生变化。我们建议毛细管阵列体模可用于纵向或多中心临床研究的质量控制。