Suppr超能文献

牙种植体植入前后的表面微观结构:一项通过扫描探针显微镜进行的体外研究

Surface microstructure of dental implants before and after insertion: an in vitro study by means of scanning probe microscopy.

作者信息

Salerno Marco, Itri Angelo, Frezzato Marco, Rebaudi Alberto

机构信息

*Research Technologist, Department of Nanophysics, Istituto Italiano di Tecnologia, Genova, Italy. †Private Practice, via Cesarea 2, Genova, Italy. ‡Technologist, Piezosurgery Inc., Columbus, OH. §Private Practice, piazza della Vittoria 8, Genova, Italy.

出版信息

Implant Dent. 2015 Jun;24(3):248-55. doi: 10.1097/ID.0000000000000244.

Abstract

INTRODUCTION

The surface microstructure of dental implants affects osseointegration, which makes their accurate topographic characterization important. We defined a procedure for evaluation of implant topography before (pre-) and after (post-) in vitro implantation test in bovine bone.

MATERIALS AND METHODS

The apical morphology of ten implants was analyzed in pre- and post-conditions using atomic force microscopy or 3D profilometry. We extracted four topographical parameters (two amplitude, 1 spatial, and 1 hybrid) and assessed the differences by analysis of variance.

RESULTS AND DISCUSSION

The implant with coating (Spline Twist MP-1 HA) was damaged. The two implants with highest pre-amplitude parameters (Pitt Easy VTPS, TLR3815) maintained their character on testing. Pitt Easy PURETEX and OT-F1 were the only nondamaged implants whose amplitude parameters increased. The surface area underwent minor changes even when the texture changed (Tri-Vent, Pitt Easy PURETEX, Exp #1). The implants that ranked the lowest in all parameters before implantation were DT4013TI, Tri-Vent, OT-F1, and Exp #2. On testing, DT4013TI showed the highest decrease in values, whereas Tri-Vent showed the highest increase in surface area. All the experimental implants showed similar topographic properties both pre- and post-test.

CONCLUSION

For most implants, no major changes occurred in surface topography on implantation. The procedure applied seems promising to evaluate the degradation of implant surface on insertion.

摘要

引言

牙种植体的表面微观结构会影响骨结合,这使得对其进行精确的形貌表征变得很重要。我们定义了一种在牛骨体外植入试验之前(预)和之后(后)评估种植体形貌的程序。

材料与方法

使用原子力显微镜或三维轮廓仪分析了10个种植体在预植入和植入后状态下的根尖形态。我们提取了四个形貌参数(两个幅度参数、一个空间参数和一个混合参数),并通过方差分析评估差异。

结果与讨论

带涂层的种植体(Spline Twist MP-1 HA)受损。两个预幅度参数最高的种植体(Pitt Easy VTPS、TLR3815)在测试中保持了其特征。Pitt Easy PURETEX和OT-F1是仅有的未受损且幅度参数增加的种植体。即使纹理发生变化(Tri-Vent、Pitt Easy PURETEX、实验1号),表面积也仅有微小变化。植入前所有参数排名最低的种植体是DT4013TI、Tri-Vent、OT-F1和实验2号。在测试中,DT4013TI的值下降幅度最大,而Tri-Vent的表面积增加幅度最大。所有实验种植体在测试前后均表现出相似的形貌特性。

结论

对于大多数种植体,植入后表面形貌没有发生重大变化。所应用的程序似乎有望评估植入时种植体表面的降解情况。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验