Suppr超能文献

SB Charité腰椎间盘假体的生物力学

Biomechanics of the SB Charité lumbar intervertebral disc endoprosthesis.

作者信息

Büttner-Janz K, Schellnack K, Zippel H

机构信息

Humboldt-Universität zu Berlin School of Medicine (Charité), German Democratic Republic.

出版信息

Int Orthop. 1989;13(3):173-6. doi: 10.1007/BF00268042.

Abstract

The SB Charité intervertebral disc prostheses consist of two metal end-plates and an interposed polyethylene slide core. These were subjected to static and dynamic testing in a servohydraulic test rig. When embedded in polypropylene, the prostheses were found to function adequately under a high static load, and under a lighter dynamic load for a long time. When tested in a cadaveric vertebral segment, a large end-plate surface area proved to be of critical importance in preventing collapse of the prosthesis into the vertebral body.

摘要

SB Charité椎间盘假体由两个金属终板和一个夹在中间的聚乙烯滑动核心组成。这些假体在伺服液压试验台上进行了静态和动态测试。当植入聚丙烯中时,发现假体在高静态负荷下以及在较轻的动态负荷下长时间都能正常发挥功能。在尸体椎体节段中进行测试时,结果表明大的终板表面积对于防止假体塌陷到椎体中至关重要。