Measurement of central corneal thickness by ultrasonic pachymeter and oculus pentacam in patients with well-controlled glaucoma: hospital-based comparative study.

作者信息

Rashid Riyam Faihan, Farhood Qasim K

机构信息

Department of Ophthalmology, College of Medicine, University of Babylon, Babylon, Iraq.

出版信息

Clin Ophthalmol. 2016 Mar 1;10:359-64. doi: 10.2147/OPTH.S96318. eCollection 2016.

Abstract

BACKGROUND

The measurement of central corneal thickness (CCT) plays an important role in the diagnosis and treatment of glaucoma and many corneal diseases.

OBJECTIVE OF THE STUDY

To compare the measurement of CCT by ultrasonic pachymeter with that measured by oculus pentacam in both normal subjects and patients with well-controlled glaucoma.

PATIENTS AND METHODS

In 173 eyes of both controls and patients with open-angle glaucoma (normal intraocular pressure) attending Ibn Al Haitham Teaching Eye Hospital, CCT was measured by oculus pentacam and then by ultrasound pachymeter at the same morning by the same technicians.

RESULTS

The results showed no significant difference in CCT readings measured by the two devices in both glaucoma and control groups (glaucoma group: 547.29±49.32 µm with pentacam vs 547.66±45.24 µm with ultrasound pachymeter; control group: 551.02±36.28 µm with pentacam vs 541.25±34.96 µm with ultrasound pachymeter). P-values were >0.05 in both groups (statistically not significant).

CONCLUSION AND RECOMMENDATION

Ultrasonic pachymeter and oculus pentacam can be used interchangeably in measuring CCT, and we recommend a nontouch method (in this study, pentacam Scheimpflug camera) for measuring CCT during assessment of patients with glaucoma or any ocular disease or surgery.

摘要

背景

中央角膜厚度(CCT)的测量在青光眼及许多角膜疾病的诊断和治疗中起着重要作用。

研究目的

比较正常受试者和青光眼病情控制良好的患者中,超声角膜厚度测量仪与Oculus Pentacam测量CCT的结果。

患者与方法

在伊本·海赛姆教学眼科医院就诊的173例对照组和开角型青光眼(眼压正常)患者的眼中,由同一名技术人员在同一天上午先用Oculus Pentacam测量CCT,然后再用超声角膜厚度测量仪测量。

结果

结果显示,在青光眼组和对照组中,两种仪器测量的CCT读数均无显著差异(青光眼组:Pentacam测量值为547.29±49.32 µm,超声角膜厚度测量仪测量值为547.66±45.24 µm;对照组:Pentacam测量值为551.02±36.28 µm,超声角膜厚度测量仪测量值为541.25±34.96 µm)。两组的P值均>0.05(无统计学意义)。

结论与建议

超声角膜厚度测量仪和Oculus Pentacam在测量CCT时可交替使用,并且我们建议在评估青光眼患者或任何眼部疾病患者或进行眼部手术时,采用非接触式方法(在本研究中为Pentacam Scheimpflug相机)测量CCT。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/8c55/4780203/f7933b6d314d/opth-10-359Fig1.jpg