VMAT 直线加速器的调试和质量保证:剂量控制和旋转架速度测试。

VMAT linear accelerator commissioning and quality assurance: dose control and gantry speed tests.

机构信息

Calvary Mater Hospital Newcastle; University of Newcastle.

出版信息

J Appl Clin Med Phys. 2016 May 8;17(3):246-261. doi: 10.1120/jacmp.v17i3.6067.

Abstract

In VMAT treatment delivery the ability of the linear accelerator (linac) to accurately control dose versus gantry angle is critical to delivering the plan correctly. A new VMAT test delivery was developed to specifically test the dose versus gantry angle with the full range of allowed gantry speeds and dose rates. The gantry-mounted IBA MatriXX with attached inclinometer was used in movie mode to measure the instantaneous relative dose versus gantry angle during the plan every 0.54 s. The results were compared to the expected relative dose at each gantry angle calculated from the plan. The same dataset was also used to compare the instantaneous gan-try speeds throughout the delivery compared to the expected gantry speeds from the plan. Measurements performed across four linacs generally show agreement between measurement and plan to within 1.5% in the constant dose rate regions and dose rate modulation within 0.1 s of the plan. Instantaneous gantry speed was measured to be within 0.11°/s of the plan (1 SD). An error in one linac was detected in that the nominal gantry speed was incorrectly calibrated. This test provides a practical method to quality-assure critical aspects of VMAT delivery including dose versus gantry angle and gantry speed control. The method can be performed with any detector that can acquire time-resolved dosimetric information that can be synchronized with a measurement of gantry angle. The test fulfils several of the aims of the recent Netherlands Commission on Radiation Dosimetry (NCS) Report 24, which provides recommendations for comprehensive VMAT quality assurance.

摘要

在 VMAT 治疗交付中,直线加速器 (linac) 精确控制剂量与旋转角度的能力对于正确交付计划至关重要。开发了一种新的 VMAT 测试交付,专门用于测试在允许的全范围旋转速度和剂量率下,剂量与旋转角度的关系。在电影模式下,安装在旋转架上的 IBA MatriXX 配备了倾斜仪,用于在计划的每个 0.54 秒测量计划期间瞬时相对剂量与旋转角度的关系。结果与从计划中计算出的每个旋转角度的预期相对剂量进行了比较。相同的数据集还用于比较整个交付过程中瞬时旋转速度与计划中的预期旋转速度。在四台直线加速器上进行的测量通常显示,在恒定剂量率区域内,测量值与计划值之间的一致性在 1.5%以内,在剂量率调制方面,与计划值相差 0.1 秒以内。瞬时旋转速度的测量值与计划值相差 0.11°/s(1 标准差)。一台直线加速器的一个错误被检测到,即名义旋转速度的校准不正确。该测试提供了一种实用的方法,可以保证 VMAT 交付的关键方面,包括剂量与旋转角度和旋转速度控制。该方法可以使用任何可以获取时间分辨剂量信息并与旋转角度测量同步的探测器进行。该测试满足了最近荷兰辐射剂量学委员会 (NCS) 报告 24 的几个目标,该报告为全面的 VMAT 质量保证提供了建议。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验