使用4D流动磁共振成像进行峰值血流速度容积评估的高效方法。

Efficient method for volumetric assessment of peak blood flow velocity using 4D flow MRI.

作者信息

Rose Michael J, Jarvis Kelly, Chowdhary Varun, Barker Alex J, Allen Bradley D, Robinson Joshua D, Markl Michael, Rigsby Cynthia K, Schnell Susanne

机构信息

Department of Medical Imaging, Ann & Robert H. Lurie Children's Hospital of Chicago, Chicago, Illinois, USA.

Department of Radiology, Feinberg School of Medicine, Northwestern University, Chicago, Illinois, USA.

出版信息

J Magn Reson Imaging. 2016 Dec;44(6):1673-1682. doi: 10.1002/jmri.25305. Epub 2016 May 18.

Abstract

PURPOSE

To test the feasibility and effectiveness of using maximum intensity plots (MIPs) based on 4D flow magnetic resonance imaging (MRI) velocity data to assess systolic peak velocities in a cohort of bicuspid aortic valve (BAV) patients.

MATERIALS AND METHODS

4D flow MRI at 1.5T was performed on 51 BAV patients. MIPs were generated from the 4D flow MRI velocity data and used by two users to determine peak velocities in three regions of interest (ROIs): ascending aorta (AAo), aortic arch, and descending aorta. 4D flow MRI peak velocities in the AAo were compared to peak velocities recorded by 2D phase contrast MRI (2D PCMRI) in a subcohort of 36 patients and by Doppler echocardiography in a subcohort of 34 patients. 4D flow MRI peak velocities recorded by each observer were compared for all ROIs to test for interobserver variability.

RESULTS

4D flow MRI recorded significantly higher velocities compared to 2D PCMRI (2.04 ± 0.71 m/s vs. 1.69 ± 0.79 m/s, 17.2% difference, P < 0.001) and similar velocities compared to Doppler echocardiography. There was excellent agreement between the observers, with a mean difference of 0.005 m/s and an intraclass correlation coefficient of 0.98.

CONCLUSION

4D flow MRI velocity MIPs allow for efficient measurement of peak velocities in BAV patients with higher accuracy than 2D PCMRI and similar accuracy to Doppler echocardiography. J. Magn. Reson. Imaging 2016;44:1673-1682.

摘要

目的

测试基于四维血流磁共振成像(MRI)速度数据的最大强度投影(MIP)评估二叶式主动脉瓣(BAV)患者收缩期峰值速度的可行性和有效性。

材料与方法

对51例BAV患者进行1.5T的四维血流MRI检查。从四维血流MRI速度数据生成MIP,并由两名使用者用于确定三个感兴趣区域(ROI)的峰值速度:升主动脉(AAo)、主动脉弓和降主动脉。在36例患者的亚组中,将AAo的四维血流MRI峰值速度与二维相位对比MRI(2D PCMRI)记录的峰值速度进行比较,并在34例患者的亚组中与多普勒超声心动图记录的峰值速度进行比较。比较所有ROI中每位观察者记录的四维血流MRI峰值速度,以测试观察者间的变异性。

结果

与二维PCMRI相比,四维血流MRI记录的速度显著更高(2.04±0.71m/s对1.69±0.79m/s,差异17.2%,P<0.001),与多普勒超声心动图记录的速度相似。观察者之间具有极好的一致性,平均差异为0.005m/s,组内相关系数为0.98。

结论

四维血流MRI速度MIP能够高效测量BAV患者的峰值速度,其准确性高于二维PCMRI,与多普勒超声心动图的准确性相似。《磁共振成像杂志》2016年;44:1673 - 1682。

相似文献

9
Improved respiratory navigator gating for thoracic 4D flow MRI.用于胸部4D流动MRI的改进型呼吸导航门控技术。
Magn Reson Imaging. 2015 Oct;33(8):992-9. doi: 10.1016/j.mri.2015.04.008. Epub 2015 May 2.

引用本文的文献

3
Cardiovascular magnetic resonance in patients with mitral valve prolapse.二尖瓣脱垂患者的心血管磁共振成像
J Cardiovasc Magn Reson. 2024 Dec 25;27(1):101137. doi: 10.1016/j.jocmr.2024.101137.
9
Multimodality Imaging in Aortic Stenosis.主动脉瓣狭窄的多模态成像
Heart Views. 2022 Jan-Mar;23(1):22-32. doi: 10.4103/heartviews.heartviews_32_22. Epub 2022 May 16.

本文引用的文献

4
Viscous energy loss in the presence of abnormal aortic flow.异常主动脉血流情况下的粘性能量损失。
Magn Reson Med. 2014 Sep;72(3):620-8. doi: 10.1002/mrm.24962. Epub 2013 Oct 2.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验